User Manual 9260
MAINTENANCE, PERFORMANCE CHECKS and ADJUSTMENTS 3-11
Pulse Response
Adjustments
The pulse response adjustments ensure that rise and fall times as
well as aberrations are within the specified range. Use this procedure
if you suspect that the pulse response is an issue.
Pulse Adjust-
ments
Equipment: Oscilloscope, PM8571A AWG
Preparation:
1. Configure the PM8571A as follows:
Function:
Standard
Waveform:
Square
Frequency:
1MHz
Amplitude:
2Vpp
Output:
On
2. Connect the 9260 input channel 1 to the PM8571A output
3. Connect the 9260 output channel 1 to the oscilloscope input
via 50Ω/2W Termination.
Adjustment:
Adjust C77 for best pulse response (<10ns, 10% aberrations)
4. Connect the 9260 input channel 2 to the PM8571A output
5. Connect the 9260 output channel 2 to the oscilloscope input
via 50Ω/2W Termination.
Adjustment:
Adjust C35 for best pulse response (<10ns, 10% aberrations)