13

Customer Service Toll-Free: 800.575.1900

Cleaning the AffloVest

Clean the AffloVest in the following situations:

  When the AffloVest becomes visibly soiled

  

Between patient uses in multi-patient care environments, 
clean the AffloVest  

  

In a single-user environment; it is recommended that the 
AffloVest be cleaned monthly 

To clean the AffloVest, please utilize a broad-spectrum 
disinfecting wipe or spray rated to kill at least 99 9% 
of bacteria and viruses and follow the manufacturer’s 
instructions for use  Using appropriate protective 
equipment, including disposable latex-free gloves, follow 
the instructions below to clean the AffloVest using the 
disinfecting wipe or spray  Follow the disinfecting wipe or 
spray label with regards to concentration, contact time, 
rinsing and drying after cleaning 
1   Ensure the AffloVest is unplugged from all power sources 
2    Remove the battery from the battery pouch and remove 

the controller from the controller pouch 

3   Unbuckle front buckles and shoulder snaps and lay 

AffloVest flat so that the inside of AffloVest is exposed 

  a   Wipe the outer edges of the AffloVest, starting in the 

collar area, tracing the outer edges all the way around 
the AffloVest and repeat 

  b   Wipe all exposed fabric thoroughly, pressing into 

creased areas, and repeat  

4   Turn AffloVest over and lay it flat 
  a       Starting at the top, wipe the exposed AffloVest fabric 

thoroughly, pressing into creased areas, and repeat 

  b   Wipe all surfaces of snaps and straps 
  c       Wipe inside and outside of buckle and buckle 

connector 

  d      Wipe inside and outside of battery pouch, including 

the plug protruding from the vest  

  e   Wipe inside and outside of controller pouch 
5   Wipe battery thoroughly, including the battery cable 
6   Wipe controller thoroughly, including controller cable 
7   Wipe all accessories thoroughly including all cables 
8   Wipe carry case thoroughly (inside and outside) 

Preventative Inspection,  
Calibration and Maintenance

On a periodic basis, the AffloVest should be cleaned as 
described in the Cleaning the AffloVest section  After 
cleaning, it is recommended to inspect the AffloVest for 
any damage to the product, wiring, battery, accessories 
and the controller  This may be performed by the user or 
by authorized service personnel  If any damage is noted, 
please contact Tactile Medical for assistance 

The AffloVest is designed to operate without any periodic 
calibration or maintenance required  Only Tactile Medical 
may replace parts or perform repair on the AffloVest 

Summary of Contents for AFFLOVEST

Page 1: ...AffloVest Instructions for Use and User Manual Frees patients Mobilizes secretions Advances therapy...

Page 2: ...Minneapolis MN 55416 USA Toll Free Tel 800 575 1900 Toll Free Fax 866 569 1912 afflovest com Welcome to AffloVest Dear Customer Congratulations on your purchase of an AffloVest from Tactile Medical W...

Page 3: ...t High Intensity Do not tumble dry Manufacturer Medium Intensity Do not iron Reference Number Low Intensity Do not wet clean Lot Number Do not machine wash Do not bleach Serial Number Waste electrical...

Page 4: ...MODEL 8052 OUTPUT 14 8VDC 5 2Ah typ 77Wh Li ion CHARGING VOLTAGE 18VDC 2 75A min Manufacturer Tactile Medical 3701 Wayzata Blvd Suite 300 Minneapolis MN 55416 USA Toll Free Tel 800 575 1900 Toll Free...

Page 5: ...and a doctor s assessment of the individual patient s case if the following conditions exist Intracranial pressure ICP greater than 20 mm Hg Recent spinal surgery or acute spinal injury Bronchopleura...

Page 6: ...us present in the air pathways in compliance with a physician s prescription and orders Application The AffloVest may be prescribed for one of these conditions Bronchiectasis Cystic fibrosis Late effe...

Page 7: ...upply plug into the AffloVest battery There is a small opening at the bottom of the battery pouch Power supply connected to battery IMPORTANT When the power supply is plugged into the battery pay clos...

Page 8: ...ttery comes connected inside of the battery pouch The battery should not require disconnection or removal from the AffloVest during normal use The serial number tag is included inside the battery pouc...

Page 9: ...treatment whether the vest is running or not P Percussion mode all motors operate in a pulsed fashion V Vibration mode all motors operate continuously D Drainage mode motors operate in a pre programm...

Page 10: ...to alert the user to charge the battery The battery for the AffloVest contains the following indicators to provide the user with useful information Fuel Gauge Displays the battery level when the butto...

Page 11: ...Treatment Time 5 15 Minutes Treatment Intensity Low Medium High Select Save and press the Play Pause button to save the settings To exit without saving select Cancel and press the Play Pause button T...

Page 12: ...the following Keep the AffloVest dry Protect the AffloVest from direct sunlight Remove the battery from the AffloVest during transport storage Keep the AffloVest between 20 C without relative humidity...

Page 13: ...racing the outer edges all the way around the AffloVest and repeat b Wipe all exposed fabric thoroughly pressing into creased areas and repeat 4 Turn AffloVest over and lay it flat a Starting at the t...

Page 14: ...problems Do not attempt to open the battery case The electrolytes and electrolyte vapors contained within the case of Lithium ion batteries may be harmful to humans If the battery case breaks or is co...

Page 15: ...eaning with any type of liquid disinfectant or cleaning solution Never put the AffloVest cords over sharp edges Never allow cords on the AffloVest to kink or bind Never carry or lift the AffloVest by...

Page 16: ...ely deposit them at municipal collection points for recyclable materials where they will be accepted free of charge Correct disposal of products will help protect the environment and avoid harmful eff...

Page 17: ...d to the AffloVest correctly Attempt to operate the AffloVest when connected to the power supply Contact your dealer for further assistance The selected program function has suddenly switched off The...

Page 18: ...ry From the AffloVest Wait sixty 60 seconds and reconnect battery The battery will beep twice when the battery is plugged back in Open battery compartment and disconnect battery as shown leave disconn...

Page 19: ...tic material the relative humidity should be at least 30 Fast Electrical Transients Bursts as per IEC 61000 4 4 2 kV for Power Supply Lines 1 kV for Input Output Lines 2 kV for Power Supply Lines 1 kV...

Page 20: ...ncy range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF interference as per IEC 61000 4 3 3 V m 80MHz to 2 5 GHz 3 V m NOTE 1 The higher frequency...

Page 21: ...DISTANCE DEPENDENT ON TRANSMIT FREQUENCY M 150 kHz to 80 MHz d 1 2 m P 80 MHz to 800 MHz d 1 2 m P 800 MHz to 2 5 GHz d 2 3 m P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100...

Page 22: ...AFFLOVEST IN NO EVENT SHALL TACTILE MEDICAL BE LIABLE FOR ANY SPECIAL DIRECT INDIRECT OR CONSEQUENTIAL DAMAGES Some states provinces or countries do not allow exclusion or limitation of incidental or...

Page 23: ...chasers are those original end user Purchasers who first took possession of the warranted item in the United States of America its possessions or territories for use of that item in the United States...