1. FOCUS STACKING 

With Focus stacking, you can stack images with different focuses on 

top of each other to create one ultra-sharp image. 

1. Magnification level

Set the magnification level of choice by using arrows up or down. 

The actual magnification level is displayed inside the zoom option and 

can also be typed if a keyboard is connected.

Plus/minus buttons on the control box are active as an alternative to 

using the onscreen zoom menu point.

2. Focus 1: Focus setting far

Manually set focus to the part of the object farthest away.

3. Focus 2: Focus setting near

Manually set focus to the nearest part of the object.

4. Steps

Choose number of images to stack between 5 and 50. 

 

Available intervals: 5, 10, 15, 20, 30, 40, 50

5. Quality

 

• 

Normal: This is the default fast mode with a typical  

 

processing speed of 8 seconds for 15 images. 

• 

High: In this mode the processing speed is reduced and   

15 images can take up to 13 seconds. 

By extending the image processing time, visual artifacts are reduced 

and the image quality increased.

6. Start

Press Start to start the process. Wait until the menu displays Done.

7. Save image

Choose this option to save the image. The image save settings will 

adapt from setup chosen from the general camera settings. 

To access saved files, see manual section ACCESS SAVED PHOTOS.

8. Exit

Close application.

NB: Long pressing the center key on the control box will close the 

Focus stacking application or cancel an ongoing focus stacking.

All current settings used during Focus stacking, except zoom level, 

are persisted during power down.

2. RULER

With Ruler, you can add vertical and/or horizontal rulers in a variety of 

colors on top of your live image. 

Get started with the app

Before using the application, make sure to specify which lens is 

currently mounted on the microscope in the microscope camera 

settings.  

If the mounted lens is not available in the microscope camera 

settings, please perform a ruler calibration. For more information, see 

section 4 in SETUP MENU. 

Ruler direction

Turn the Ruler application on in the Applications menu with the left 

and right arrow buttons on the XPLUS control box. Choose between 

Horizontal, Vertical or Both.

Turn off

Turn the ruler off by once again accessing Ruler option under the 

Applications menu point and choose Off.

System units

Choose your preferred system units (metric or imperial) under the 

General settings menu point in the setup menu.

Ruler color

Set the ruler color to your prefered color to separate it from your 

displayed object. Go to the Ruler color option under the General 

settings menu point in the setup menu.

Choose between 6 different colors: Yellow, blue, red, black, white and 

green. 

 

Manual focus setting

When the Ruler is chosen, the system will automatically start in 

manual focus setting to focus on the part of the object furthest away. 

If needed, manually adjust the focus by adjusting the height of the 

microscope to get the desired object in focus. Then, use the zoom in 

and zoom out buttons to fine adjust the focus.

9.  STANDARD APPS

STANDARD APPS (1/2)

Summary of Contents for FHD PRESTIGE

Page 1: ...SETUP MENU 10 9 STANDARD APPS 15 11 SYSTEM UPDATE 18 12 LICENSE ACTIVATION 20 13 TECHNICAL SPECIFICATIONS 21 14 PC REQUIREMENTS 23 15 MAINTENANCE 24 16 WARRANTY 24 18 ERGONOMIC RECOMMENDATIONS 24 19...

Page 2: ...ways turn off the system before unplugging when possible WARNINGS The product is a digital magnifying system consisting of a camera unit PCBs mechanical parts and a power supply The product is intende...

Page 3: ...ffusor lens kit Fresnel lens kit 100 mm for Lens 1x Plan 4 YOU HAVE RECEIVED Smart guide Camera arm with lens Base Rod washers screw OPTIONS XPLUS FHD control box Power supply USB A B 3 0 cable Hex ke...

Page 4: ...lly and horizontally on top of the base Fasten the two washers with the screw on top of the rod Pull the release button and place the camera arm on the rod The complete microscope with base 5 ASSEMBLI...

Page 5: ...trol box Output 3 3V 6A Max Use only control boxes supplied by TAGARNO 4 Output HDMI Standard Type A Monitor HDMI Out 1080p60 5 Ethernet RJ 45 Ethernet wall socket Ethernet LAN 100BASE TX 1000BASE T 6...

Page 6: ...Power on off Place the camera arm in the appropriate height according to the lens you are using Adjust the friction according to your personal preference 7 OPERATION TIP Find the How to Use tutorial v...

Page 7: ...ou wish to focus on something very close Push this button on the control box to adjust focus if you wish to focus on something that is farther away To turn the autofocus on again simply push both zoom...

Page 8: ...to USB or your microscope internal storage press the image capture button until you hear a beep Long press on the same button will save an image with graphics causing the live image to freeze for a fe...

Page 9: ...e right and left arrows to choose between the different settings in the microscope menu Activate an application or submenu by pressing Enter If you have selected Exit pressing Enter will close down th...

Page 10: ...lat surface Follow the on screen directions to find the optimal working distance The magnification settings prior to performance will be restored 3 Close up Lens Indicate which lens is mounted on your...

Page 11: ...haracters missing in the scanned image name 16 Image preview This feature allows users to quickly inspect and capture images of new samples When an image is captured the default behavior is to preview...

Page 12: ...Japanese Korean Polish Romanian Russian and Spanish 11 Keyboard layout Select your horizontally language by pressing buttons left or right on the XPLUS control box The available languages are listed a...

Page 13: ...top of your live image on screen For more info see manual section STANDARD APPS 4 Watermark Create an info field on captured images with an image prefix time stamp horizontal field of view microscope...

Page 14: ...torage does not currently have this file system use this function to format the USB storage A second pop up will ask you to confirm your selection Please notice that all files on the USB memory stick...

Page 15: ...or cancel an ongoing focus stacking All current settings used during Focus stacking except zoom level are persisted during power down 2 RULER With Ruler you can add vertical and or horizontal rulers i...

Page 16: ...ings at the same time 2 Enable or disable info field Enable or disable an info field automatically stating an image prefix default or as stated under Camera settings time stamp horizontal field of vie...

Page 17: ...ss of method verify that an IP address of the microscope is displayed in the microscope System information If no IP address is stated check that all cables are connected correctly or consult a network...

Page 18: ...k www tagarno com firmware download 2 Download the installation file s to a USB memory stick FAT32 formatted 3 If the file s is in a ZIP folder unzip the files and make sure that only the correct inst...

Page 19: ...m update STEP 5 VERIFY INSTALLATION 1 Verify the system update by checking that the Application version displayed in the Setup menu Info System information matches the version of the system update fil...

Page 20: ...it with a unique license key file To activate a license key on your TAGARNO microscope follow the steps below Note The license can only be paired with the microscope which has the serial number state...

Page 21: ...Monitor Detect Yes Plug connection 2 x USB 2 0 Type A Socket USB 3 0 output Type B Socket Control box D SUB 9 Pin Female ESD Snap Socket 10mm Monitor output HDMI Standard Type A Power supply DC Jack E...

Page 22: ...tor arm Flexarm XPLUS FHD Control box Table mount monitor arm Cleaning kit Foot switch Wall mount monitor arm Dust cover TAGARNO Ring light White Circular tilting table Coaxial light TAGARNO Ring ligh...

Page 23: ...Vidia GeForce GT 740M with 2GB RAM Monitor resolution 1920x1080 FULL HD 1080P 25 30HZ OS Desktop Laptop Windows 7 8 or 10 with DirectX 11 Memory 4GB CPU Intel Core Dual 3 0GHz or i3 2 4GHz 4th generat...

Page 24: ...will be free from defects in material and workmanship for a period of 24 months 2 years from date of invoice from Supplier This warranty covers a The replacement of defective parts b All labour costs...

Page 25: ...PTION TAGARNO A S hereby declares that the product listed above consisting of a camera unit and a 12V power supply is in compliance with the following European directives 2006 25 EU 2014 30 EU 2014 35...

Page 26: ...ncy energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to...

Reviews: