background image

Part no.

Product name

Spec

Color

Designer

K100

311-4141100-041

機器說明書/NO LOGO/EN(BG+CHOL+UA)/.標準品/   

Emily

L148*W105mm/

模造紙

/

40

/80P/

/

騎馬釘

  

Summary of Contents for E1

Page 1: ...Part no Product name Spec Color Designer K100 311 4141100 041 NO LOGO EN BG CHOL UA Emily L148 W105mm 40 80P...

Page 2: ...E1 Multi Functional Monitoring System Owner s Manual...

Page 3: ...Version 4 0 2021 08 311 4141100 041...

Page 4: ...Control Tests 13 Prepare For Blood Tests 15 Perform Blood Tests 16 Blood Glucose Tests 18 Total Cholesterol Tests 20 Uric Acid Tests 22 Review Test Results 24 Data Transfer via Cable 25 Data Transfer...

Page 5: ...diseases Please refer to the table below for what kind of whole blood sample to obtain Home Use Professional Use Blood Glucose Capillary Capillary Venous Arterial Neonatal Total Cholesterol Capillary...

Page 6: ...proximity to sources of strong electromagnetic radiation as these may interfere with the accurate operation 8 Proper maintenance and control solution tests are essential to the longevity of your devi...

Page 7: ...used for the testing of newborns total cholesterol and uric acid test strips must NOT be used for the testing of newborns 15 If your biochemical parameter results are lower or higher than usual and yo...

Page 8: ...est strip by pushing up this button 2 Display Screen 3 Down Button 4 Up Button 5 Test Strip Slot Strip Port Communication Insert test strip here to turn the meter on for testing Connect strip port cab...

Page 9: ...time of day AC before meal PC after meal Auto QC Mode QC control solution test 2 Face Low High Symbol 3 Memory Mode 4 Code 5 Measurement Unit 6 Date Time 7 Low Battery Symbol 8 Total Cholesterol Symb...

Page 10: ...de of the test strip should face up when inserting into the test strip slot Test results might be inaccurate if the contact bar is not fully inserted into the test strip slot The E1 Multi Functional M...

Page 11: ...10 Letter Indicator Guide g a j k l b c d e f h i G lu c o s e U ri c A c id C h o le s te ro l Refer to this letter indicator guide when necessary...

Page 12: ...e press to select the correct number Press 2 Setting the time format and time Press to select the desired time format 12h or 24h Press With the HOUR MINUTE flashing in sequence press to select the cor...

Page 13: ...MINUTE flashing in sequence press to select the correct HOUR MINUTE Press and go to the next alarm setting NOTE When the alarm beeps press to switch it off Otherwise it will beep for 2 minutes then sw...

Page 14: ...ntrol solution test you will need b c and d 1 Insert the test strip to turn on the meter Insert the test strip into the meter Wait for the meter to display a flashing and GLU CHOL UA 2 Apply the contr...

Page 15: ...l solution beyond the expiration date or 3 months after first opening Write the opening date on the control solution vial and discard the remaining solution after 3 months It is recommended that the c...

Page 16: ...ons in the test strip insert Prepare the Lancing Device for Blood Tests Please follow the instructions in the lancing device insert on how to prepare the lancing device and collecting a blood sample W...

Page 17: ...against the lower side of your fingertip Press the release button to prick your finger A click indicates that the puncture is complete 4 Obtaining a blood sample i Gently squeeze the punctured site to...

Page 18: ...you have trouble filling the confirmation window please contact your health care professional or customer service for assistance 6 Read Your Result The result of your test will appear after the meter...

Page 19: ...rm study showed that maintaining blood glucose levels close to normal can reduce the risk of diabetes complications by up to 60 1 The results provided by this system can help you and your healthcare p...

Page 20: ...lt is between 130 to 239 mg dL 7 2 to 13 2 mmol L Result is between 180 to 239 mg dL 10 to 13 2 mmol L Result is between 120 to 239 mg dL 6 6 to 13 2 mmol L Result is 240 mg dL 13 3 mmol L Ketone Warn...

Page 21: ...are displayed in milligrams of total cholesterol per deciliter of blood mg dL or in millimoles of total cholesterol per liter of blood mmol L Total Cholesterol mg dL or mmol L Desirable 200 mg dL 5 1...

Page 22: ...sult is 100 mg dL 2 5 mmol L Out of measurement range Result is between100 to 400 mg dL 2 5 to 10 3 mmol L Result is 400 mg dL 10 3 mmol L Out of measurement range Please consult your doctor to determ...

Page 23: ...lasma can either have a weight dimension mg dL or a molarity mol L The approximate calculation rule for conversion of mg dL in mol L is mg dL Times 59 48 mol L E g 10 mg dL x 59 48 594 8 mol L mol L D...

Page 24: ...S Result is 3 mg dL 178 mol L Out of measurement range Result is between 3 to 20 mg dL 179 to 1189 mol L Result is 20 mg dL 1190 mol L Out of measurement range Please consult your doctor to determine...

Page 25: ...and hold and at the same time to switch off the meter Reviewing Blood Glucose Average Results 1 Start with the meter switched off Press and release to enter memory mode for average results with displa...

Page 26: ...t System please visit the TAIDOC s website www taidoc com 2 Connecting to a personal computer Connect the strip port cable to a cable port on your computer With the meter switched off insert the other...

Page 27: ...of your current condition and to achieve better diabetes control Pairing with your mobile device 1 Turn on the Bluetooth function on your mobile device 2 Take off the cap and insert the Bluetooth don...

Page 28: ...ing on your device is turned on and the meter is within the receiving range before transmitting the data Please find OS version requirement on App Store or Google Play when you download the app The Bl...

Page 29: ...treated as a contaminated item that may carry a risk of infection during measurement The batteries in the used meter should be removed and the meter should be disposed in accordance with local regulat...

Page 30: ...en the symbol appears along with display messages The meter is functioning properly and the result remains accurate but it is time to change the batteries 2 When the symbol appears with E b The power...

Page 31: ...from children If swallowed seek medical assistance immediately battery might leak chemicals if left unused for a long time Remove the battery if you are not going to use the device for an extended per...

Page 32: ...date shown on the packaging If the problem persists repeat with a new code chip E t Ambient temperature is above or below system operation range Repeat the test after the meter and test strip are wit...

Page 33: ...y Test strip inserted upside down or incompletely Insert the test strip with contact bars end first and facing up Defective meter or test strips Please contact customer services 2 If the test does not...

Page 34: ...nd repeat the test again Expired or contaminated control solution Check the expiry date of the control solution Control solution that is too warm or too cold Control solution meter and test strips sho...

Page 35: ...tion Use by date Humidity limitation Batch code CE mark Serial number RoHS compliance Catalogue number This device does not belong to household waste and must be returned to a collection point for rec...

Page 36: ...ning Test Strip Operating Conditions Blood Glucose 8 C to 45 C 46 4 F to 113 F and 10 to 90 R H non condensing Total Cholesterol Uric Acid 10 C to 40 C 50 F to 104 F and 10 to 85 R H non condensing Me...

Page 37: ...to 10 3 mmol L Uric Acid Test 3 to 20 mg dL 178 to 1189 mol L Expected service life 5 years Operating Altitude Up to 2000m for indoor use Degree of Pollution Pollution degree 2 This device has been te...

Page 38: ......

Page 39: ......

Page 40: ......

Page 41: ...TaiDoc Technology Corporation B1 7F No 127 Wugong 2nd Rd Wugu Dist 24888 New Taipei City Taiwan www taidoc com MedNet EC REP GmbH Borkstra e 10 48163 M nster Germany For self testing 0123 TD 4141...

Reviews: