background image

4. Replace and twist the cap to close 

the lock as shown in the picture until you 
hear the click. 
When the lock is properly closed lock 
indicator is in the extreme right position.

Hold your fingers on the
body NOT the cap. 

1

2

3

INSTRUCTIONS FOR USING 

Setting up the Lancing Device

Alternative Site Testing (AST)

Cleaning

x

The Lancing Device allows you to collect your capillary blood from fingertip (and 

alternative sites) easily and virtually pain free.

2. Insert a lancet into the lancet holder 
and push down firmly until it is fully 
secured.

3.  Twist the protective disk off the lancet.

7. Lancing the finger

Press the lancing device’s tip firmly 

against the lower side of your fingertip. 

Press the release button to prick your 

finger, then a click indicates that the 

puncture is complete.

8. Remove the device from the puncture 

site. After penetration, remove the first 

drop of blood with a clean cotton swab. 

Gently squeeze the punctured area to 

obtain another drop of blood. Be careful 

NOT to smear the blood sample. Apply 

the blood to the test strip.

10. Slide the ejector forward to remove 
the used lancet. Dispose of the lancet 
according to your local regulations.
 ?  

For collecting blood samples from 

sites other than the fingertip, use the 
clear cap.

Testing on sites other than the fingertip is referred  

Please read the 

Alternative Site Testing

 section in your  

you perform AST. 

Substitute the lancing device cap with the clear cap 

alternative site testing. Pull the cocking 

When lancing the forearm, upper arm, hand, thigh, or calf, avoid lancing the areas 
with obvious veins in order to prevent excess bleeding.

9. Remove the lancet by twisting the 
cap off first. Safely dispose of the used 
lancet by placing the protective disk on a 
hard surface, and pushing the exposed 
tip into the disk. 

5. Holding the cap select on the depth of 
penetration by turning the adjustable tip 
in either direction so that the arrow on the 
cap points to the desired depth.

6. Pull the cocking control back until it 
clicks. You will see a color change inside 
the release button when it is ready. 

?  

If it does not click, the device may 

have been cocked when the lancet 

Referent

was inserted.

The Lancing Device offers you 6 depths of skin penetration:

Symbol

The longer the length of the indicator line, the greater the depth of penetration.

Shallow

Deep

1

2

1

2

1. Twist and pull off the cap of the 
lancing device.

To reduce the chance of infection:

?  Never share a lancet or the lancing device.
?  Always use a new, sterile lancet. Lancets are for single use only.
?  Avoid getting hand lotion, oils, dirt or debris in or on the lancets and the lancing 
  device.
?  The used lancet may potentially be biohazardous. Discard it according to your   

WARNING

    local regulations.

Before collecting a drop of blood:

?  Wash and dry your hands before you start.
?  Select the puncture site at the fingertip.
?  Rub the puncture site for about 20 seconds before penetration.

NOTE

?  Clean the site using cotton swab moistened with 70% alcohol and let it air dry.

Community 

Sterilized using irradiation 

Do not use if package is damaged

STERILE R

Ver 1.0  2014-12

318-5010200-xxx

Cap

Cocking
Control

Release
Button

Lancet Holder

Ejector

Arrow

Depth Indicator

Adjustable Tip

A Clear Cap
(For AST use only)

Lock Indicator

Lock

The lancet is disposable and is intended 

single use only. The lancing device is reusable.

To clean the lancing device, use a mild  soap and

water to wipe clean the lancing device.

DO NOT rinse the device under water. Disinfect

the cap by placing it in 70% alcohol for 10 minutes

and then allow it to air dry. Do NOT place the device

in a dishwasher or use detergents.         

for

Soft or thin skin

Average skin

Thick or calloused skin

Marketed and Distributed by:

Rapid Check Diagnostics & Wellness Ltd.

11, Balogun Street, Off Obafemi Awolowo Way,
Ikeja, Lagos, Nigeria. P.O Box 499 Ikeja
Tel: 234-1- 892 9914; +234- 803 411 2811; +234- 807 189 1575
Email: info@rapidcheckng.com, sales@rapidcheckng.com
website: www.rapidcheckng.com

KIPTRACK LANCING DEVICE

to as alternative site testing (AST). 

Owner’s  Manual  before

for 

control back until it clicks.  

Symbol Information

Do not reuse

Consult instructions before use

Use by / Expiry date

Symbol   Referent

Batch code

Manufacturer

Caution, consult accompanying documents

Authorized representative in the European

TaiDoc Technology Corporation

B1-7F., No. 127, Wugong 2nd Road, Wugu Dist.,
New Taipei City 248, Taiwan (R.O.C)

MedNet GmbH

Borkstrasse 10, 48163 Muenster Germany

EC REP

Reviews: