background image

 

Dear TD-4283 System Owner: 

 
Thank you for purchasing the

 TD-4283 

Blood Glucose Monitoring System. 

This manual provides important information to help you to use the system 
properly. Before using this product, please read the following contents 
thoroughly and carefully.   

Regular monitoring of your blood glucose levels can help you and your doctor 
gain better control of your diabetes. Due to its compact size and easy 
operation, you can use the 

TD-4283 

Blood Glucose Monitoring System to 

easily monitor your blood glucose levels by yourself anywhere, any time. 
 
If you have other questions regarding this product, please contact the place of 
purchase or call the Customer Care Line.   
 

 
 
 
 
 
 

 

Summary of Contents for TD-4283

Page 1: ...TD 4283 BLOOD GLUCOSE MONITORING SYSTEM OWNER S Manual Version 1 0 2012 02...

Page 2: ...nts thoroughly and carefully Regular monitoring of your blood glucose levels can help you and your doctor gain better control of your diabetes Due to its compact size and easy operation you can use th...

Page 3: ...onsult your doctor to have the results interpreted 7 Before using this device to test blood glucose read all instructions thoroughly and practice the test Carry out all the quality control checks as d...

Page 4: ...OD SAMPLE Preparing the Puncture Site Alternative Site Testing Performing a Blood Glucose Test METER MEMORY Reviewing Test Results Reviewing Blood Glucose Day Average Results Downloading Results onto...

Page 5: ...are experiencing symptoms that are inconsistent with your blood glucose test results and you have followed all the instructions given in this owner s manual contact your healthcare professional We do...

Page 6: ...ould not be used for the diagnosis of diabetes or testing on newborns Professionals may test with capillary and venous blood sample home use is limited to capillary whole blood testing Test Principle...

Page 7: ...meter on for testing 2 Display Screen 3 Main Button Enter the meter memory and silence a reminder alarm 4 Battery Compartment 5 Set Button Enter and confirm the meter settings 6 USB Port only for char...

Page 8: ...Display Screen...

Page 9: ...the test strip into the slot Contact Bars Insert this end of the test strip into the meter Push it in firmly until it will go no further ATTENTION The front side of test of test strip should face up w...

Page 10: ...he Date With the year flashing press Main Button until the correct year appears Press SET With the month flashing press Main Button until the correct month appears Press SET With the day flashing pres...

Page 11: ...hing press Main Button to add an hour Press SET to confirm and go to minutes press Main Button to add one minute Hold Main Button longer to add faster Press SET to confirm and go to the next alarm set...

Page 12: ...hout regard to time since last meal AC no food intake for at least 8 hours PC 2 hours after a meal QC testing with the control solution You can switch between each mode by 1 Start with the meter switc...

Page 13: ...control solution test when you first receive the meter at least once a week to routinely check the meter and test strips you begin using a new vial of test strips you suspect the meter or test strips...

Page 14: ...horoughly before use Squeeze out a drop and wipe it off then squeeze out another drop and place it on the tip of the vial cap Hold the meter to move the absorbent hole of the test strip to touch the d...

Page 15: ...king properly Do NOT test your blood Contact your local customer service or place of purchase for help NOTE The control solution range printed on the test strip vial is for control solution use only I...

Page 16: ...the glucose value obtained Blood from a site that has not been rubbed exhibits a measurably different glucose concentration than blood from the finger When the puncture site was rubbed prior to blood...

Page 17: ...T Pull the cocking control back until it clicks When lancing the forearm upper arm hand thigh or calf avoid lancing the areas with obvious veins because of the risk of excessive bleeding NOTE Choose a...

Page 18: ...nges faster than capillary blood at other sites Thus when testing blood glucose during or immediately after a meal physical exercise or any other event take a blood sample from your finger only We str...

Page 19: ...r to the FOUR MEASURING MODES section 3 Obtaining a blood sample Use the pre set lancing device to puncture the desired site Wipe off the first appeared drop of blood with a clean cotton swab The size...

Page 20: ...turn off You must remove and reinsert the test strip to start a new test The confirmation window should be filled with blood before the meter begins to count down NEVER try to add more blood to the te...

Page 21: ...ntainer The meter will turn itself off automatically after the test strip is removed Always follow the instructions in the lancing device insert when removing the lancet WARNING The used lancet and te...

Page 22: ...s 1 Press and release Main Button will appear on the display Press Main Button again and the first reading you see is the last blood glucose result along with date time and the measurement mode 300 mg...

Page 23: ...tored in each measuring mode in the order of Gen AC then PC 3 Exit the meter memory Keep pressing Main Button and the meter will turn off after displaying the last test result NOTE Any time you wish t...

Page 24: ...and installing the software To download the Health Care Software System please visit the TaiDoc s website www taidoc com 2 Connecting to a personal computer With the monitor turned off the meter will...

Page 25: ...he USB cable to the USB port of the meter 2 Connect the other end of cable to a free USB port on your personal computer 3 The charging indicator will light up in red when battery is recharging After t...

Page 26: ...6o F to 89 6 below 85 relative humidity Do NOT freeze Store your test strips in their original vial only Do not transfer to another container Store test strip packages in a cool dry place Keep away f...

Page 27: ...nths It is recommended that the control solution test be done at room temperature 20 C to 25 C 68 F to 77 F Make sure your control solution meter and test strips are at this specified temperature rang...

Page 28: ...to repair the meter yourself and never try to disassemble it under any circumstances Result Readings MESSAGE WHAT IT MEANS 20 mg dL 1 1 mmol L 20 69 mg dL 1 1 and 3 8 mmol L 70 129 mg dL 3 9 7 2mmol L...

Page 29: ...the system s operating range You cannot perform a test until the meter and test strip are within the operating range of 10 C to 40 C 50 F to 104 F Repeat the test after the meter and test strip are w...

Page 30: ...switch off 2 minutes after last user action Repeat the test with a new test strip Apply sample only when flashing appears on the display Defective meter Please contact customer services 3 If the cont...

Page 31: ...for people without diabetes mg dL Fasting and before meal 100 mg dL 5 6 mmol L 2 hours after meals 140 mg dL 7 8 mmol L Source American Diabetes Association 2010 Clinical Practice Recommendations Dia...

Page 32: ...p away from sunlight Keep dry Temperature limitation Use by Expiry date Use within three months after first opening Batch code Manufacturer Serial number Caution consult accompanying documents Authori...

Page 33: ...rtion Automatic reaction time count down Automatic blood detection Temperature warning Operating condition 10 C to 40 C 50 F to 104 F below 85 R H Storage transport conditions 20 C to 60 C 4 F to 140...

Page 34: ...television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate...