28
Preparations for Operation
OM8-T53
Focus switch
for descent
Focus switch
for rise
Stroke mark
RED
REFLEX
OFF
㪉㪃㪌
㪈
㪇㪃㪍 㪇㪃㪋
㪈
㪈 㪃㪍
・In case of using Type Ⅱ Foot switch unit
1.Step on the right or the left of the foot switch
to rise or descent the microscope so that the
center mark on the microscope comes to the
center of the stroke mark.
OM8-T54
X-Y centering switch
・In case of using X-Y coupling unit
1.Press the X-Y centering switch on the X-Y
coupling unit .
・The X-Y coupling unit will return to its original
position (the center of its range of movement).
■ Adjusting Diopters
● Prior to operation, adjust the diopters of the microscope's eyepieces to your eyesight.
● If you wear glasses, fold back the eyecaps of the eyepieces and then adjust the diopters.
● For adjusting the diopters, have clean white paper ready.
1.Press the [ I ] side of the power switch, turning
on the power. Turn the microscope light
intensity control knob clockwise, turning on the
light.
2.Place white paper below the objective lens, turn
the diopter adjustment ring of then eyepieces
to be adjusted in the counterclockwise
direction (+ side) until it stops, and look into the
eyepieces.
・The contours of the visual field look unclear.
3.Turn the diopter adjustment ring back in the
clockwise direction (- side) until the contours of
the visual field become clear.
4.Repeat Steps 2 to 3 for the other eye.
OM8-T55
White Paper
Diopter adjustment ring
+
−