Calibration
196
Offset/Full Scale Calibration of Output Current Monitor (Isolated Output Current
Monitor)
1. Disconnect the load wires and the remote sensing
wires. Connect the digital multimeter to the output
terminal.
Turn the POWER switch ON and OUTPUT OFF in
advance.
Power Supply
+
-
V
+
-
Digital
Multimeter
Calibration Circuit
V
+
-
Digital
Multimeter
1
14
External
Control Connector
Insulation
Control Connector
2
1
2. Set the parameter for FUNCTION item 12 to "0" to
disable Linearity Compensation Function.
On how to set, refer to Linearity Compensation
Function under Useful Functions (page58).
3. Set the constant voltage setting value to [ xV ] and
the constant current setting value to [ yA ] and set
the output to ON.
30-200
30-400
60-100
60-200
[xV]
1.00V
1.00V
1.00V
1.00V
[yA]
0.0A
0.0A
0.0A
0.0A
500-12
500-24
1000-6
1000-12
[xV]
1.00V
1.00V
1.00V
1.00V
[yA]
0.00A
0.00A
0.000A
0.00A
120-50
120-100
400-15
400-30
[xV]
1.00V
1.00V
1.00V
1.00V
[yA]
0.00A
0.0A
0.00A
0.00A
On how to set, see Operating as Constant Voltage
Power Supply under Basic Operations. (page
4. Press the FUNC. key.
5. Turn the VOLTAGE dial to display the setting item
number 34 in the Number Display (voltage).
* At that time, turn off the Number Display (current).
6. Adjust so that the voltage monitor value becomes
"0.00V" with the CURRENT dial.
7. Press the FUNC. key to settle the offset calibration
value of output voltage monitor (isolated output
current monitor).
Summary of Contents for HX-S-G2 Series
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...