Over Current Protection (OCP)
In the event of an over current due to a short circuit in the load, the output is turned OFF and switching is ceased to
protect the load.
The OCP current can be set from 1% to 110% of the rated current.
Setting/Canceling Over Current Protection
Setting Over Current Protection Level
. Set the output to OFF with the OUTPUT key.
. Press the OVP/OCP key to display the current OCP set value in the Numbe
r Display (current).
. Change the set value by turning the CURRENT dial while pressing down the OVP/OCP key.
. Release the OVP/OCP key to confirm set values and to return to measurement display mode.
Holding down
Verifying Over Current Protection Operation
After settings are complete, please follow the procedures below to test verify operation.
. Disconnect the load from the power supply and short between the +/- output terminals using cross section area wires
that can flow rated current.
. Turn the POWER switch ON with the Output ON/OFF Toggle mode (FUNC.51) set to "0".
. Set the Output Current settings lower than OCP level.
. Turn OUTPUT to ON.
. Gradually increase the Output Current with the CURRENT dial. Verify
that the output turns OFF once the OCP set
current has been reached and that the error code "E005" appears in the Number Display (voltage) and the Number
Display (current) becomes the "OCP" display.
. Hold the OVP and OCP keys simultaneously for more tha
n 1 second to cancel the alarm. Return the CC set value to
its original value.
● Press the CURRENT dial to display the current number
position. The selected number will be brightened. Press
to change position.
If all 4 places are brightened, numbers will change from
the lowest place.
● Press the OVP/OCP key to display the OVP set value in
the Number Display (voltage) and the OCP set value in
the Number Display (current).
● If the number of parallel devices (total power) is changed,
the OCP set value is set again to the maximum value.
MEMO
* The figure above is an example of
connecting (HX-S-030-200G2)
Summary of Contents for HX-S-G2 Series
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...
Page 239: ......