Using via Digital Data Communication
94
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
: Sets output voltage
Function:
Sets the output voltage
Format:
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRf>
Type:
SET command, QUERY command
Parameter:
30VType: 0.00 to 31.50/ 60VType: 0.00 to 63.00/ 500VType: 0.0 to 525/ 1000VType: 0 to 1050
Response:
Same as parameters
Restrictions:
Alarm response is returned for normal SET commands in the following conditions
·
While operating as a series operation slave unit
·
While operating as a parallel operation slave unit
<Ex.>
VOLT 30
VOLT?
30.00
[SOURce]:VOLTage:PROTection[:LEVel]
: Sets OVP voltage
Function:
Sets the OVP voltage
Format:
[SOURce]:VOLTage:PROTection[:LEVel] <NRf>
Type:
SET command, QUERY command
Parameter:
30VType: 0.30 to 33.00/ 60VType: 0.60 to 66.00/ 500VType: 5.0 to 550.0/ 1000VType: 1 to 1100
Response:
Same as parameters
Restrictions:
Alarm response is returned for normal SET commands in the following conditions
·
While operating as a parallel operation slave unit
<Ex.>
VOLT:PROT 33
VOLT:PROT?
33.00
Summary of Contents for HX-S-G4 Series
Page 157: ...Special Loads 157 This page is blank...
Page 161: ...Maintenance 161 This page is blank...
Page 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Page 198: ...Specifications 198 2 HX S 030 400G4 F I...
Page 199: ...Specifications 199 3 HX S 060 100G4 F I...
Page 200: ...Specifications 200 4 HX S 060 200G4 F I...
Page 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Page 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Page 203: ...Specifications 203 7 HX S 01000 6G4 F I...