Calibration
178
Output Current Control with External Resistance (B)
1. Perform load with only the shunt resistor.
Disconnect the load wires and the remote sensing
wires.
Connect the shunt to the output terminal. Connect
the digital multimeter to the terminal located on the
shunt.
Turn the POWER switch ON and OUTPUT OFF in
advance.
Note: Large current flows through
the wiring material.
Use wires which will not produce
a voltage drop below 0.8V.
Calibration Circuit
Power Supply
+
-
+
-
V
6
19
External
Control Connector
External
Resistance
0 to 10k
Ω
2. Set the constant voltage setting value to [ xV ]
and the external resistance to 0Ω and set the
output to ON.
30-200
30-400
60-100
60-200
[xV]
1.00V
1.00V
1.00V
1.00V
500-12
500-24
1000-6
1000-12
[xV]
1.0V
1.0V
1V
1V
For how to set, refer to Output Current Control with
External Resistance (B).(page145)
3. Press the FUNC. key.
4. Turn the VOLTAGE dial to display the setting item
number 30 in the Number Display (voltage).
* At that time, turn off the Number Display (current).
5. Adjust so that the output current becomes "0A"
with the CURRENT dial.
6. Press the FUNC. key to settle the offset calibration
value of output current control with external
resistance (B).
Summary of Contents for HX-S-G4 Series
Page 157: ...Special Loads 157 This page is blank...
Page 161: ...Maintenance 161 This page is blank...
Page 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Page 198: ...Specifications 198 2 HX S 030 400G4 F I...
Page 199: ...Specifications 199 3 HX S 060 100G4 F I...
Page 200: ...Specifications 200 4 HX S 060 200G4 F I...
Page 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Page 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Page 203: ...Specifications 203 7 HX S 01000 6G4 F I...