Basic Usage
45
Error Codes
Place
Items
No. Display
(Voltage)
No. Display
(Current)
Content
Verify/Recovery
WDT
E001
-
CPU may have
performed an illegal
operation
Turn POWER switch OFF and restart
OHP
E003
OHP
Over heating protection
is triggered.
Turn power OFF. Leave device to cool
down. Turn power ON
Check environmental temperatures
Ve
rify Fan Motor rotates when power is
turned on.
OVP
E004
OUP
Over voltage protection
was triggered
Check OVP values.
See page 40 about how to cancel.
OCP
E005
OCP
Over current protection
was triggered
Check OCP values
See page 42 about how to cancel.
BST
E006
-
Slave unit
malfunctioned.
Check slave unit display.
P_ERR
E007
-
May have had a short
interruption or voltage
dip at the main power.
May be damaged.
Contact Takasago Sales Dept.
BST_NRDY
E008
-
Slave unit's input
voltage is abnormal.
Check whether slave unit’s power switch is
ON
Phase
interruption
detection alarm
Phase interruption of
input power is
detected.
Please check the wiring connection to
the input connector.
Device Detection
Error
E009
-
Cannot detect number
of parallel devices
Restart power supply
No. of Parallel
Connected
Devices
Over
E010
-
More than 11 parallel
devices are connected
Verify No. of parallel devices. Restart power
supply
Device No.
Disagreement
E011
-
Difference in the
current No. of
connected parallel
devices and the
previous No. of parallel
devices remembered
by the power supply
If the present number of parallel devices is
satisfactory, hold the FUNC. key for more
than 2 seconds and confirm the detected
number of devices.
TRIP
E012
Shdn
TRIP function was
triggered
Cancel by pressing OVP+OCP keys
simultaneously or by the remote control
command ALM REG CLEAR.
Models of
different voltages
mixed
E013
-
Models of different
voltages are
connected.
Verify connection with models of different
voltage and restart the power supply.
Communication
Error
E100
-
Data was not received
Re-verify sent message.
Recovered if normal message is received.
E2P Error
E110
-
Failure to initialize
during startup
Cancel by pressing OVP+OCP keys
simultaneously or by the remote control
command ALM REG CLEAR.
Summary of Contents for HX-S-G4 Series
Page 157: ...Special Loads 157 This page is blank...
Page 161: ...Maintenance 161 This page is blank...
Page 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Page 198: ...Specifications 198 2 HX S 030 400G4 F I...
Page 199: ...Specifications 199 3 HX S 060 100G4 F I...
Page 200: ...Specifications 200 4 HX S 060 200G4 F I...
Page 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Page 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Page 203: ...Specifications 203 7 HX S 01000 6G4 F I...