Using via Digital Data Communication
------------------------------------------------------------------------------------
81
OUTPut Level
OUTPut level command corresponds to functions concerning output control.
Command Name
Summary
Command Type
OUTPut
-
-
:DELay
-
-
:ON
Sets Output ON Delay Time
Both
:OFF
Sets Output OFF Delay Time
Both
:EXTernal
-
-
:MODE
Sets operation mode for output control at External Contact
Both
:HOT
Sets output status when POWER is ON
Both
:MODE
Sets Startup Mode
Both
[:STATe]
Sets Output ON/OFF
Both
OUTPut:DELay:ON
:Sets Output ON Delay Time
Function: Sets Output ON Delay Time
Same as FUNCTION Settings Item 8
Format:OUTPut:DELay:ON <NRf>
Type:SET command, QUERY command
Parameters: 0.00~99.99
Response: Same as parameters
Restriction: Alarm response is returned for normal SET commands in the following conditions
・While operating as a series operation slave unit
・While operating as a parallel operation slave unit
・Sequence operation in progress
<Ex.>
OUTP:DEL:ON 12.34
OUTP:DEL:ON?
12.34