Controlling with External Analog & Contact Signals
------------------------------------------------------------------------------------
121
Uses of Terminals
① Toggling Output ON/OFF at External Contacts
② Cut off input with external contacts
③ Control output voltage/current with external voltage
④ Control output voltage/current with external resistance
⑤ Output monitor
⑥ Output status
⑦ Master-Slave Series Operation
How to Insert Wire into Connector
① Push a flat-head screwdriver into the
square hole.
② Insert the stripped wire (as shown in
the right diagram) in the hole while the
screwdriver is inserted.
※Spring Connection
※No pin is required for connecting
the wire.
Flat-Head Screwdriver Dimensions
MEMO
Wire Specifications
・Wire Size:AWG 28-18
・Stripped Length:7mm
0.4mm
2mm
7mm
マイナスドライバー
マイナスドライバー
線材
Wire
Flat-Head Screwdriver
Flat-Head Screwdriver
③Extract Screwdriver