Basic Operations
------------------------------------------------------------------------------------
27
Setting Output ON/OFF Toggle Mode
It is possible to nullify the use of the OUTPUT key.
To do so, set the settings to 1, and when the POWER switch is turned ON, output will begin even if the OUTPUT
key is not pressed.
Setting Procedures
Press the Function FUNC key.
“FUNC.” Lamp lights.
The Setting Item Number will display on the top Number Display
and the parameter will be displayed in the bottom Number Display.
Select Item No. 51 with the VOLTAGE dial.
Set Parameter to 0 or 1 with the CURRENT dial.
Parameter=0 Output is OFF (COLD START)after power
is turned on
Pressing the OUTPUT key will toggle it
ON/OFF.
Parameter=1 Output turns ON approximately 2 seconds after
power is turned on.
(HOT START)
Press the FUNC. key to finalize settings.
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Summary of Contents for ZX-S-H Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Page 37: ...Basic Operations 33 This is blank page ...
Page 59: ...Useful Function 55 This is blank page ...
Page 115: ...Using via Digital Data Communication 111 This is blank page ...
Page 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Page 147: ...Special Loads 143 This is blank page 守 ...
Page 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Page 189: ...Appendix 185 ZX S 800H 800HA ...
Page 190: ...Appendix 186 ZX S 1600H 1600HA ...