The Incubating Rocking/Waving Shakers have been designed for the temperature,
speed/tilt angle and time functions to work independently of one another. The
temperature and speed/tilt angle can be reset without resetting the timer and the
timer can be stopped and started without interrupting the heating or rocking/waving
functions.
1. getting ready:
a. Plug the power cord into a properly grounded outlet. The standby indicator
light will illuminate, verifying power to the unit. Manually move the tray to a near
horizontal “home” position before powering unit.
b. Press the standby button to move the unit from standby mode. The standby
indicator light will shut off and the temperature, speed/tilt angle and time displays
will illuminate, displaying the previously used settings.
2. Setting temperature:
a. Press the up/down arrows below the temperature display until you reach the
desired temperature. When you release the arrow button, the display will blink
off and then on indicating the new set temperature has been accepted.
b. Press the on/off button to start the heating function. The indicator light below the
temperature display will illuminate to indicate the heating function is in use. You
will hear five (5) audible beeps that indicate set temperature has been reached.
c. Temperature adjustments can be made without interrupting heating by using
the up/down arrows below the temperature display. After the change has been
made and you release the button, the display will blink off and then on indicating
the new set temperature has been accepted.
d. To stop the heating function, press the on/off button below the temperature
display. The heat indicator light will turn off.
P
rofessional
i
nCubating
r
oCking
/W
aving
s
haker
o
Perating
i
nstruCtions
Caution Hot indicator:
The caution hot indicator light warns that the temperature of the
air in the chamber is above 40°C (104°F). The light will illuminate
and remain lit when the temperature of the air in the chamber
reaches approximately 40°C (104°F). When the heat is turned
off, the caution hot indicator light will stay lit until the temperature of the air in the
chamber is less than 40°C (104°F).
3. Setting tilt angle: Electronic tilt enables the user to adjust the angle of tray
movement while the unit is rocking/waving or while the unit is stopped.
a. Press the function button below the speed/tilt angle display until the tilt indicator
light illuminates. You are now ready to set the tilt angle.
b. Press the up/down arrows below the speed/tilt angle display until you reach
the desired angle. When you release the button, the display will blink off and
then on indicating the new set angle has been accepted. The unit will complete
one rotation at the previously set angle, then smoothly change to the newly
programmed angle. The tilt indicator light will flash until the transition to the new
tilt angle is complete.
4. Setting speed:
a. Press the function button below the speed/tilt angle display until the speed
indicator light illuminates. You are now ready to set the speed.
b. Press the up/down arrows below the speed/tilt angle display until you reach the
desired speed. When you release the button, the display will blink off and then
on indicating the new set speed has been accepted.
c. Press the on/off button to start the rocking/waving function. The indicator light
below the speed/tilt angle display will illuminate and blink until the set-point is
reached. Once the set-point is reached the light will stop blinking and remain lit
until rocking/waving has ceased.
6