Van ’t Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
E-mail: sales@tamson.com
Website: www.tamson.com
Tamson Instruments bv
Page 12/20
tc16-40-58-man.docx Rev. 2.4 UK 9-21
ISO 9001 : 2015
NL/PRO 238239125
VAT: NL 80 66 34 984 B01
Bank account no.:
NL28 INGB 0007 350 370
NL95 RABO 0160100046
Chamber of commerce 27 16 95 41
7.3
Backside of the thermostatic bath
1
Cooling fan
2
RS 232 Communication
3
In and outlet pump
side view connection:
4
Mains cable
5
Motor fuse
6
Cooling coil, 1/4" outher thread
7
Switch for fan (only for TC16 with p/n 31T0671/
31T0861)
8
Fan (only for TC16 with p/n 31T0671/31T0861)
7.4
Selecting parameters
The front panel layout shows the following four operating
keys:
Press shortly and following is displayed
SP.1 - Temperature set point in °C,
tU.n
-
Tuning the bath (“Atune”),
Pb
- Changing the tuning (PID) parameters "P"value
ti
- Changing the tuning (PID) parameters "I"value
td
- Changing the tuning (PID) parameters "D"value
OFSt
- Change temperature readout offset (± 5°C)
Each time is pressed another option is displayed.
Up and Down keys allow changing the listed value. All
changed values, like set point and PID parameters, will
be kept in memory when pressing the page key to leave
2
1
3
7
4
6
5
8
PAGE
DOWN
UP
LIST