9.2
Message on display
Code
What does it mean
What action to take
E61
Defect / calibration PT100
Check wiring and element. Ohmic value
at 0
C=100R and at 100
C=138R.
Replace PT100
E70
Error on operational parameters
E50
EEprom error
Auto - reset after 2 s
E82
Error on input during calibration
E4
EEprom error
Auto - reset after 2 s
E5
EEprom error
Auto - reset after 2 s
E6
Watchdog error
Auto - reset after 2 s
10 SPECIFICATION
10.1 Technical specifications overview
Item
Unit
TC16
TC40
TC58
Ordering
code
230V
50 or
60Hz
00T0671
00T0681
00T0691
115V
50 or
60Hz
00T0861
00T0851
00T0881
Power
[Watt]
1500
2900
2900
Materials
Used inside bath: stainless steel 304, brass
Range
Ambient .. 250
C/..482
F
Reading
Standard
C,
F on request
Setting ±
[°]
0.1
Stability ±
[°K]
0.02
0.02
0.02
Heating
[W]
1400**
2800
2800
Bath
volume
[L]
16
40
58
Pump
[mBar]
Max pressure 300*
Pump
[L/min]
Max flow 7.5*
Opening
[mm]
180 *210
420 *275
420 *275
Depth
bath
[mm]
220
200
300
Length
[mm]
455
705
705
Width
[mm]
295
375
375
Height
[mm]
440
440
590
Weight
[Kg]
21
30
35
CE
Conforms to CE regulation
* Optional pump with pressure max 1bar, flow max
16ltrs/min, reference
:
24T0392. See specification sheet
"pumpgraph
** Additional heater 1400W optional to heat larger loads