The cause for this relatively high temperature is the self 
heating of the bath due to friction-heat of the pump. The 
insulation  of  the  bath  prevents  this  energy  to  be 
transferred into the environment. This causes a raise of 
temperature in the bath up to 25 .. 30°C depending on the 
environmental temperature and used bath fluid. 
  
For  cooling  the  bath  down  to  lower  temperatures,  tap-
water  is  mostly  used.    Also  a  small  circulator  can  be 
connected to the cooling coil like the Tamson TLC10-3, 
or TLC15. 
 

6.4 

Over-temperature safety 

 
In case of electronic failure the possibility exists that the 
heater  element  is  continuously  switched  on.  This  will 
cause  extreme  temperature  raise.  To  prevent  high 
temperatures  the  bath  is  fitted  with  a  mechanical 
temperature  switch  (thermostat).  This  thermostat  will 
switch-off the bath at a preset temperature in the range 
from 50 tot 270°C. 
 
We advise to adjust the mechanical over temperature to 
approxi 25 C above the bath set point.  
 
This  safety  construction  prevents  for  example  oil  to  be 
heated above flash point which definetly will cause fire or 
prevent  evaporation  of  bath  fluid  due  to  high 
temperatures. 
 
The  thermostat  will  automatically  reset  when  the  bath 
temperature drops approximately 10°C below the pre-set 
temperature. To continue normal operation the bath has 
to be switched-off and on again. 
 

6.5 

Adjusting 

The 

Mechanical 

Over-

Temperature Protection Thermostat 

 
• 

Turn the over-temperature protection (8, drawing 

2) to its maximum (turn clockwise towards 50°C), 
• 

Heat the bath to its set point, 

• 

Gently turn the over-temperature protection anti-

clockwise,  until  the  over-temperature  protection  is 
activated, and system switches-off, 
• 

Turn 

the 

over-temperature 

protection 

approximately 25°C higher (turn clockwise), 
Switch  the  bath  off  and  on  again.  The  bath  is  ready  to 
operate safely. 
 
 
 

 

Summary of Contents for 00T0671

Page 1: ...il sales tamson com Website www tamson com Tamson Instruments bv Page 1 20 tc16 40 58 man docx Rev 2 4 UK 9 21 ISO 9001 2015 NL PRO 238239125 VAT NL 80 66 34 984 B01 Bank account no NL28 INGB 0007 350 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 USER MANUAL TC16 40 58 ...

Page 2: ...CTION TO THE TC SERIES 8 6 1 GENERAL 8 6 2 PUMP 8 6 3 COOLING 8 6 4 OVER TEMPERATURE SAFETY 9 6 5 ADJUSTING THE MECHANICAL OVER TEMPERATURE PROTECTION THERMOSTAT 9 6 6 FLUID LEVEL 10 6 7 FLUID LEVEL DETECT 10 7 OPERATION 11 7 1 SWITCHING ON 11 7 2 CONTROL PANEL 11 7 3 BACKSIDE OF THE THERMOSTATIC BATH 12 7 4 SELECTING PARAMETERS 12 8 QUICK START 14 8 1 TUNING THE BATH 14 8 2 TUNE PID VALUES 15 8 3...

Page 3: ... damage to the equipment or its accessories caused by incorrect operation Important When performing service maintenance or moving the apparatus always disconnect the apparatus at the main s socket Proper skilled and trained personnel are only allowed to operate this equipment Take notice of warning labels and never remove them Refer service and repairs to qualified technician If a problem persists...

Page 4: ...nel Avoid damage to the equipment or its accessories through incorrect operation Environment Panel sealing Confirms EN60529 IP65 Environment Temperature 0 tot 35 C Supply enough ventilation Humidity 5 tot 95 non condensating Atmosphere Not suited for altitudes above 2000m or explosive corrosive environment Pollution cat 2 Conducting pollution must be prevented 4 INSTALLATION 4 1 Important Tamson I...

Page 5: ...cate included in the box as well as ASTM thermometers thermometer holders etc Please see the packing list for details concerning total contents of consignment Before filling the bath remove any remaining packing material from its interior The interior of the bath can be accessed by taking off the lid on the top of the bath 4 3 Initial use Before filling the bath make sure that the bath is clean Pl...

Page 6: ...ating temperature When viscosity of the bath fluid is too high at room temperature the motor fuse will be activated In such a case the bath fluid can first be heated to set point temperature When the bath has reached it s set point the motor fuse can be reset by pressing it This way of operating will reduce the life of the motor and is thus not recommended Type Remarks Life Viscosity mm s X Temp R...

Page 7: ... 370 NL95 RABO 0160100046 Chamber of commerce 27 16 95 41 Do not use demineralized or distilled water This water can lead to corrosion of the bath and moving parts The corrosion can wear out the pump bearings 5 CONNECTING Before plugging TC 16 TC40 or TC58 into the mains socket make sure the voltage of the bath corresponds to the local voltage and frequency Use a mains supply that is well earthed ...

Page 8: ...hrough a closed external system The standard pump offers a motor protection small thermal fuse which is placed on the back panel of the motor compartment This protects the motor from excessive loads When pumping viscous liquids the thermal fuse can be activated The motor fuse can be reset by pressing the button The inlet and outlet of the pump are fitted with 8 mm hose connections and 1 4 threaded...

Page 9: ...ll cause extreme temperature raise To prevent high temperatures the bath is fitted with a mechanical temperature switch thermostat This thermostat will switch off the bath at a preset temperature in the range from 50 tot 270 C We advise to adjust the mechanical over temperature to approximately 25 C above the bath set point This safety construction prevents for example oil to be heated above flash...

Page 10: ...s too low When the bath has been installed it must be filled with an appropriate liquid When working with water the bath should be filled to 1 cm below the lid For oil the bath should be filled to not more than 5 cm below the lid Depending on the operating temperature the liquid level in the bath should be observed and excessive fluid should be removed The liquid level should be maintained between...

Page 11: ...ng on If the bath has been properly filled with fluid it can be switched with the mains switch located on the front panel Choose a working temperature set point 7 2 Control Panel 1 LED Display 2 LED Over temperature 3 LED Heater 4 Up key 5 Down key 6 Page key 7 Scroll key 8 Safety thermostat 9 Mains On Off 10 Top lid When the overtemperature safety is activated wait until the bath temperature is l...

Page 12: ... 7 Switch for fan only for TC16 with p n 31T0671 31T0861 8 Fan only for TC16 with p n 31T0671 31T0861 7 4 Selecting parameters The front panel layout shows the following four operating keys Press shortly and following is displayed SP 1 Temperature set point in C tU n Tuning the bath Atune Pb Changing the tuning PID parameters P value ti Changing the tuning PID parameters I value td Changing the tu...

Page 13: ...ID values can not be set manually anymore pb Changing the tuning PID parameters Proportional band value Default value is 1 range 1 100 Proportional band equals 1 proportional value Press page key to confirm setting when altered ti Changing the tuning PID parameters Integrator value Default value is 200 Range 1 1200 Press page key to confirm setting when altered td Changing the tuning PID parameter...

Page 14: ...eir own heat capacity the use of external cooling and external connected processes circulation or working at different set point temperatures requires new settings of the PID parameters These parameters have to be optimized after changes to the system when optimal and accurate temperature control of the bath liquid is required Tuning of the bath results in Stable temperature control of the bath No...

Page 15: ...below provides fast finding of the PID settings and is refereed to as the method of Ziegler Nichols This procedure is only to be followed when automatic tuning is not functioning This method provides fast manual findings of the PID values Start the bath at its required running temperature 1 Set the integral time Ti and the derivative Time Td to off 2 Check if the Lcb and Hcb are set to auto 3 Igno...

Page 16: ...ent temperature to remove the fluid make sure that the hose to drain and waste container can withstand the temperature of the bath fluid When removing bath fluid do not inhale toxic vapor Always use appropriate ventilation 8 6 Using the drain tap optional The TC baths can be drained via the drain tap located at the backside of the apparatus The drain has an inner diameter of 3 8 gas fitting The dr...

Page 17: ...Instruments bv Problems with set point Heater LED is not burning motor is turning and temperature raises above set point Setpoint too near to room temperature Cooling of the bath is needed Temperature doesn t reach set point motor turns fast Bath fluid evaporates too quick Use other fluid Heater malfunctions Measure electric mains current Current below 2 amps indicate a problem with the heating el...

Page 18: ...m error Auto reset after 2 s E6 Watchdog error Auto reset after 2 s 10 SPECIFICATION 10 1 Technical specifications overview Item Unit TC16 TC40 TC58 Ordering code 230V 50 or 60Hz 00T0671 00T0681 00T0691 115V 50 or 60Hz 00T0861 00T0851 00T0881 Power Watt 1500 2900 2900 Materials Used inside bath stainless steel 304 brass Range Ambient 250 C 482 F Reading Standard C F on request Setting 0 1 Stabilit...

Page 19: ...d safe handling procedures should be provided to handlers and users Alteration of this document is strictly prohibited Except to the extent required by law republication or retransmission of this document in whole or in part is not permitted Tamson Instruments bv assume no responsibility for accuracy of information unless the document is the most current available Tamson Instruments neither repres...

Page 20: ...06 42 EC Machinery Directive of the European Parliament and of the Council of 17 May 2006 42 EC 2nd Edition June 2010 p d EN 60204 Machinery Directive and Safety requirements p pi e EN60950 1 Low Voltage Directive p f EN61000 3 2 Harmonics p g EN61000 3 3 Flicker p3 h EN61000 4 2 A1 A2 ESD p i EN61000 4 3 A1 A2 Radiated immunity p anechoic room j EN61000 4 4 Electrical Fast Transients Minimum requ...

Reviews: