- 21 -
7.1.7 Bath
Level
The bath liquid level is low.
Replenish the bath liquid.
7.1.8 Bath
Overheat
The bath temperature has exceeded the set temperature of the bath overheat cutoff
device.
Set the temperature of the bath overheat cutoff device to the test tempe 10
°
C or
higher.
7.1.9 Replace
Battery
The memory backup battery is flat.
Replace the battery.
7.2 Troubleshooting without Error message
Phenomenon
Cause and Check Item
Solution
Circuit protector
head projecting
Replace the CPU
board.head to reset.
Unable to start
Circuit protector
head not projecting
Replace the MAIN
board.
No display
Able to start
Faulty display
Replace the
fluorescent display.
Able to start
Replace the
fluorescent display.
Abnormal display
Unable to start
Replace the MAIN
board.
Intermittent buzzer
sounds after MAINS
switch ON
Faulty CPU board
Replace the CPU
board.
Other keys also disabled
Replace the MAIN
board.
Operation key disabled
Other keys not disabled
Replace the panel
switch sheet.
Heater wire broken
Replace the heater.
Bath heating failure
Heater wire not broken
Replace the start
relay or the heater
SSR.
Bad result in Counter
Check
Crystal oscillation in CPU is defective
Replace the CPU
board.
Stirring failure
Faulty stirrer motor
Replace the stirrer
motor.