9
3. Operation
3.1 Power ON
Turn the MAIN switch ON. A message screen as illustrated in
Fig. 4
appears on the control Unit
display.
Fig. 4 Main Screen
3.2 Entering a Barometric Pressure Value (not necessary when built-in pressure sensor is
equipped)
APM-7 is capable of making automatic correction for barometric pressure by entering the
barometric pressure reading through the control panel. Enter the barometric pressure before a flash
point test starts as follows:
(1)
Press
FUNCTION
key, The display shows
Baro set
screen as illustrated in
Fig. 5
appears. (If
the temperature unit is selected
o
F, (mmHg) will appear instead of (hPa).
(2)
Press
Left
or
Right
key to appear the cursor under the barometric pressure value.
(3)
Enter the barometric pressure reading that was obtained from a calibrated barometer using the
four arrow keys (
Left-Right, Up-Down
keys).
Fig. 5 Barometric Pressure Entering Screen
APM TEMP EXP.FP
Electric
270
-A-
o
C
23.4
Model Type
Ignition Source (appear when electric configuration only)
Expected
Flash Point
Procedure & Mode
Specimen Temperature
APM Baro set(hPa)
1013
Prev.FP 137.0