12
· IN Pos err Optional
Enabled when the sensor position detection function is
added
Temp
Sensing temperature of temperature sensor
Photo
→
L R
Left and right receiver input levels. When LAMP is set
to ON, the input levels are automatically adjusted to
about 50%.
3. Adjustment
3.1 Ring Stage Adjustment
To avoid detection errors, inspect the position of the ring stage periodically. If mounting nuts
are loose, check the mounting dimensions of the sample shelf and bottom plate. Adjust the
dimensions shown in Fig. 13 and tighten the nuts.
From the top of the ring holder
to the lower face of the top plate
: 100+/-0.1mm
100
From the lower face of the
Ring holder to the upper face
of the bottom plate: 30+/-0.1mm
30
Fig. 13
Dimensions of the Ring Holder
According to ASTM D36, the bottom of the shouldered rings in the ring holder shall be 25
mm above the upper surface of the bottom plate. ASP-5 is using photoelectric detection
system. If the light beam is passed just above the bottom plate, reflected beam from the
bottom plate disturb the correct detection. The axis of light beam of ASP-5 is passing 5 mm
above the upper surface of the bottom plate. So, ASP-5 detects ball falls 25 mm below the
bottom of the shouldered rings.