6
· Det.sens
The default is 85%. A drop is detected when the receiver input level has
dropped to "Average detector input over preceding 10 to 20 seconds
×
Detection sensitivity
÷
100".
During a test, check the status on the secondary display (press the
CHECK key).
(See Fig.7)
Status
Projector output
Receiving Level (L-R)
MEAS
P_OUT RATE REC LEVEL
00%
0.0
°
C
50% 49%
00%
25.0
°
C
35% 34%
H_OUT
TEMP DET LEVEL
Heater output
Sample drop detection level
Fig. 7 Sub Screen
· Add.nenu
Two additional test modes can be added to the original two test
modes.
To complete testing a sample, select On here and also at "1-sample"
On the measuring range setting screen.
In addition, the safety stop temperature can be changed.
· Rounding
Usually set to On. Rounds the average test result to the nearest
0.5
°
C.
1.2.4 External connection settings
Printer, PC, and other external connections can be set as follows:
Ext.out
Inst No.
Test No.
Computer
01
0001
Opt.prt
Sample ID
Off
Off
End
Fig. 8
External Out Settings Screen
Secondary
display