1.3 Temperature Unit Selection and Temperature Correction
The temperature correction screen is used to make any corrections if there are any
differences between the temperature display and the softening point thermometer.
The tester handles two kinds of bath liquid (water and glycerin). Compare the temperature
sensor (displayed temperature) of the tester and the softening point thermometer while
using the necessary bath liquid in advance to check the differences.
1.3.1 Procedure of Calibration
Insert the thermometer suitable for the selected liquid bath into the specified hole in the ring
stage and secure it with an elastic band or wire at the standard position. After checking that
the temperature sensor is also at the specified position, place a beaker containing the bath
liquid at the specified position of the measuring unit, put the ring holder in the beaker, and
attach the beaker to the ring holder.
Select the bath liquid using the MODE key and start a test. Compare and record the
thermometer reading and temperature display every 10
°
C. (Copy the attached settings table
for recording data during this operation.)
Repeat the above operation twice and pigeonhole data as references for correction.
1.3.2 Select Bath Medium and Temperature Unit.
Select Temp Cal from Menu screen, Fig 9 appears first. Select Water or Glycerin (blinks)
、
then press FUNCTION key. Fig.10 appears by this operation. Push Right or Left key to blink
“
°
C”, and then change it by Up or Down key. If you want to go Temperature offset entering
screen, select Next, if you want to finish, select End and then push FUNCTION key.
Fig.9
Bath Liquid Selection Screen Fig.10
Bath Liquid Selection Screen
Correct differences at each correction point, which are at every 20
°
C from 0 to 200
°
C on the
following temperature correction screen for each bath liquid:
Example
Correction when the 40
°
C reading is 0.5
°
C lower than the
thermometer reading
In Fig. 10, the current temperature correction value for 40
°
C is
−
1.0
°
C and the
necessary correction value is +0.5
°
C.
New correction value
=
-1.0
°
C
+0.5
°
C
=
-0.5
°
C
→
-0.5 for 40
°
C
Bath select
Water
Glycerin
End
Set temp unit End
°
C
Next