Part Names
2. Connecting an Input Module
Insert the module into the module jack. Once the Unit recognizes the module, the LCD display will change as shown
below and recording will start. (If you have purchased an RTR-505-P, the Unit has been ready for the pulse measure-
ment by default, therefore recording will start upon installation of the battery.
The factory default settings are as follows: Recording Interval at 10 minutes, Recording Start at Immediate Start, Recording Mode at Endless.
Make sure that the module is completely
inserted until you hear a “click” sound.
Set Model Number Measurement Items Input Module
LCD Displayed Items (see the following section #3)
Compatible Sensors
RTR-505-TC
Temperature
Thermocouple Module (TCM-3010)
Measurement, Unit of Measurement, Sensor Type,
Operational Status
Thermocouple: Type
K, J, T, S
RTR-505-Pt
Temperature
PT Module (PTM-3010)
Measurement, Unit of Measurement, Sensor Type,
Operational Status
Pt100, Pt1000
RTR-505-V
Voltage
Voltage Module (VIM-3010)
Measurement, Unit of Measurement, Operational Status
Voltage Input Signal
RTR-505-mA
4-20mA
4-20mA Module (AIM-3010)
Measurement, Unit of Measurement, Operational Status
4-20mA Input Signal
RTR-505-P
Pulse
Pulse Input Cable (PIC-3150)
Measurement, Unit of Measurement, Operational Status
- - - -
1. Installing the Battery
Cover
Rubber
Packing
Lithium Battery
(inserted into tube)
Screws
1
Remove the screws and open the cover.
Make sure to use the proper size and type of screwdriver. (Phillips head #1 screwdriver is recommended.)
2
Insert the included battery.
Do not remove the battery from its tube casing.
- If using a CR2 lithium battery, the tube is not necessary.
3
Check the rubber packing for any cuts or scratches and close the cover as it was when
opened.
- Dust or defects on the packing can adversely affect the waterproof capacity; in this case, remove the dust or replace the pack-
ing if it is damaged.
- Be sure to completely close the cover.
- Make sure not to over tighten the screws.
(Appropriate Tightening Torque: 20N cm to 30N cm{2Kgf cm to 3Kgf cm})
Notes about Battery Installation
- After inserting the battery for the fi rst time, nothing may appear or occur for about 10 seconds; this is not a malfunction.
- If a new battery has been installed and nothing appears in the display, please remove and reinsert the battery.
- When inserting a battery, make sure no water or foreign objects get inside the case.
- Make sure that + and - are in the correct direction.
About Lithium Batteries
- Lithium batteries CR2 sold in stores may also be used, but if you are using in an environment below -20°C, above 60°C, or in a
situation such as transportation where continued vibration is likely to occur, we strongly suggest the purchase and use of the our
lithium battery LS14250. (Please purchase the optional battery set TR-11P2 in which LS14250 is included.)
- When using a LS14250 type lithium battery, even though a new battery has been inserted the battery warning mark may remain
on for a short time. This is due to a special characteristic of the battery. Note that the longer the battery has been in storage the
longer the amount of time, from 10 minutes to about 1 hour, the battery warning mark will remain on. If during that time the Base
Unit is used to get the current status of the Remote Unit, the remaining battery level will show that the battery level is low.
- Please store the lithium battery LS14250 in a place that is 20 °C or less.
- When changing a CR2 battery, we strongly suggest changing the rubber packing and the drying agent. Please purchase the
maintenance set TR-00P1 sold separately.
RTR-505 is a Remote Unit. In order to use wireless
communication,it is necessary to purchase a Base Unit
separately. (Compatible Base Units: RTR-500, RTR-500NW,
RTR-500AW)
Wireless Data Recorder RTR-505 is referred to as the “Unit”
in this manual.
Specifications
Device Name
RTR-505
Measurement Items (*1)
Temperature / Voltage / 4-20mA / Pulse
Recording Intervals
Select from 15 choices: 1, 2, 5, 10, 15, 20, and 30 sec. / 1, 2, 5, 10,
15, 20, 30, and 60 min.
Logging Capacity
16,000 Readings
Recording Start Method
Immediate Start / Programmed Start
Recording Mode
Endless / One Time (Factory Default Setting: Endless)
LCD Displayed Items
Measurement, Recording Status, Recording Mode, Battery Life
Warning, Unit of Measurement, Full (Logging Capacity FULL),
Sensor Unconnected, Input Module Unconnected, Measurement
Range Exceeded, Display Range Exceeded)
Communication
Interfaces
Wireless Communication / Optical Communication
Wireless Communication
Specifi cations
RF Power
7mW
Radio Standard
Specifi cations
FCC Part15 Section247 / IC RSS-210
Frequency Range : 902 to 928MHz
Transmission Range
About 150m (500ft) (if direct and unobstructed)
Communication
Speed
About 2 min. (when downloading 1 unit at full logging capacity)
Optical Communication
Speed
About 160 sec. (when downloading 1 unit at full logging capacity at
2400bps)
Power
Lithium Battery (LS14250) / Lithium Battery (CR2) (*2)
Battery Life (*3)
About 10 months (for temperature measurement)
Waterproof Capacity
Splash proof (rated for use in daily life) (*4)
Dimensions
H62
×
W47
×
D19mm (excluding protrusions and sensor part)
Weight
About 55g (including battery / excluding input module)
Operating Temperature
Range
Unit Temp Resistance: - 40 to 80
°
C
(Unit temp resistance and measurement range is -40 to 80
°
C but
wireless communication cannot occur in an environment of less
than -30
°
C.)
Other
A Base Unit is required (sold separately).
(Compatible Base Units: RTR-500, RTR-500NW, RTR-500AW)
(*1) Measurement items other than Pulse will have different measurement range depending on the input
module and sensor to be connected. For details, see the User
s Guide included with the input
module.
(*2) The included lithium battery (LS14250) is not sold in stores. Please purchase the optional battery set
for low-temperature use (TR-11P2) for replacement.
(*3) Battery life varies depending upon the connected input module, measuring environment, frequency
of communication, Unit settings, and battery performance. For details, see "Estimated Battery Life"
in this manual.
(*4) Waterproof capacity of the Unit with an input module connected. Note that sensor connection of the
input module is not water resistant.
Wireless Data Recorder
RTR-505
User’s Manual
Thank you for purchasing our product.
Carefully read this instruction manual before using this Unit.
T&D Corporation
http://www.tandd.com/
817-1 Shimadachi Matsumoto, Nagano 390-0852 JAPAN
Fax:+81-263-40-3152
© Copyright T&D Corporation. All rights reserved.
This is printed using recycled paper.
2011.05 16504760001 (1st Edition)
Outline of RTR-505
External Input Module Type
Splash proof (rated for use in daily life)
RTR-505 is a data logger designed to measure and
record different items depending on the input module to
be connected: temperature (thermocouple/Pt), voltage,
4-20mA, and pulse. The body is splash proof (rated for
use in daily life), which can be placed in an environment
between -40 and 80
°
C. Recorded data can be
automatically collected from the RTR-505 via wireless
communication with the Base Unit (sold separately), and
also viewed in graph and table form as well as printed
out by using the supplied software.
Package Contents:
Data Logger (RTR-505), Tubed Lithium Battery (LS14250),
Input Module(*), Strap, User's Manual (this manual
including warranty)
* An input module included in the package differs
depends upon which "set model" has been purchased.
3. How to Read the LCD Display
When being used in very hot or cold environments the display may become difficult to read. This is not a malfunction.
Basic LCD Display
[REC] Mark
The recording status will appear.
ON: Recording in progress
BLINKING: Waiting for programmed start
[ONETIME] Mark
When the recording mode is set to "One Time", this
mark will appear. The factory default setting is "End-
less" and this mark will not appear.
Battery Warning
Mark
When it is time for the battery to be replaced, this
mark will appear.
Measurements and
Messages Area
Measurements or operational messages will appear
here.
Sensor Type
Types of sensor connected or set to the Unit will
appear.
Thermocouple Sensor: K, J, T, S
Platinum Thermal Resistance Sensor: Pt (Pt100),
PtK (Pt1000)
Unit of Measurement
The unit of measurement for the display will appear.
Battery Replacement Sign
1.
When it is time for the battery to be replaced, a battery
warning mark will appear.
Please replace the battery as soon as possible if this
mark appears.
2.
If you do not change the battery and continue using the
Unit, the measurement display will intermittently
display [bAtt].
- After this point the downloading of data can no
longer be done via wireless communication.
- Without changing the battery and attempting to use
optical communication to download recorded data,
the communication may be broken and if so all
recorded data may be lost.
- Recording will be continued.
3.
If the battery is further left unchanged, the display will
automatically shut off.
If, at this time, a new battery is placed in the Unit,
[CHEC] will appear on the display after which record-
ing will begin again using the previously set recording
conditions. Note however that all previously recorded
data will have been lost.
Estimated Battery Life
When a battery warning mark appears, try to replace the battery with a new
one as soon as possible.
In a normal temperature environment, where recorded data is downloaded
once a day or if monitoring is carried out once every ten minutes:
Set Model Number
Recording Interval: 1
second
Recording Interval: 10
seconds or longer
RTR-505-TC
About 10 months
About 10 months
RTR-505-Pt
About 10 months
About 10 months
RTR-505-V, RTR-505-mA
About 10 months
About 10 months
RTR-505-P (Input: Open)
About 20 months
About 10 months
RTR-505-P (Input: Short)
About 20 months
About 10 months
- When the average reading recording function is enabled for the RTR-505-V or RTR-
505-mA, the battery life will be the same as when the recording interval is one
second regardless of the actual recording interval.
- The battery warning mark may appear sooner than noted above.
- Battery life will be shortened if used under the following conditions: downloading
data very often, setting the recording interval at less than ten seconds, or measuring
in an environment below -20 °C (-4 °F) or above 60 °C (140 °F).
Notes about Changing the Battery
- Before replacing a battery, please make sure to download any necessary data and
proceed with changing the battery.
- If + (plus) and – (minus) are mistaken, or if the battery ter and – are shorted,
the recorded data that is stored in the Unit will be lost.
- Downloading of data cannot occur while the battery is removed.
Display Example for Each Model
Display varies on different models as shown below.
RTR-505-TC (Thermocouple)
Temperature measurement will be displayed. (Unit:
°F / °C) Sensor type will be displayed under the
measurement; the factory default setting is Type K.
By using the software included with the Base Unit,
you can change the sensor type.
RTR-505-Pt (Pt100 / Pt1000)
Temperature measurement will be displayed. (Unit:
°F / °C) Sensor type will be displayed under the
measurement; the factory default setting is Pt100.
By using the software included with the Base Unit,
you can change the sensor type.
RTR-505-V (Voltage)
Voltage measurement will be displayed. (Unit: V /
mV) Due to the wide measurement range, the Unit
has been set by default to adjust the decimal point
automatically to display the measurement in V. By
using the software included with the Base Unit, you
can change the unit of display.
RTR-505-mA (4-20mA)
4-20mA measurement will be displayed. (Unit: mA)
Optical
Communication
Area
Module
Sensor Jack
Antenna
LCD Display