29
Basic Function - TR-71U/72U/73U
Adjustment Settings:from the [Settings] Menu
By setting adjustment values beforehand, you can record and display
the post- adjusted measurement values. You can choose from two
adjustment methods:
1
-point and
2
-point. Adjustment will be carried out
using an adjustment equation of Y
=
aX
+
b; where X is the pre-
adjusted measurement value and Y is the post-adjusted value.
Rough Guidelines for Adjustment
1
point Adjustment: Use when measuring in a range of
±
20
℃
.
2
point Adjustment: Use when measuring in a wide range.
NOTE:
- When adjusting at
2
points, make sure the difference in points is at least
10
℃
.
- When measuring in a wide range and adjusting at
2
points, make sure that the adjustment
values reflect the range you are measuring.
- We cannot guarantee that after carrying out adjustment the measuring accuracy will improve for
all measuring ranges.
1
.
Check the channel you wish to adjust.
[Read File] Button
[Save File] Button
Check here
[History] Column
Check one, and enter
the value
[Send Settings]
Button
2
.
Place a checkmark next to either [
1
Point Adjustment] or [
2
Point
Adjustment] and enter the [Before] and [After] values.
3
.
By clicking the [Send Settings] button, the adjustment settings will be
completed.
- The data logger can store up to
4
adjustment values, the adjustment history will display
up to
4
channels at one time.