TR-71nw / TR-72nw / TR-72nw-S / TR-75nw / TR-75wb - Specifications
TR-71nw
TR-72nw
TR-72nw-S
TR-75nw
TR-75wb
Measurement Channels
Temperature 2ch
Temperature 1ch,
Humidity 1ch
Temperature 1ch,
Humidity 1ch
Temperature 2ch
Sensor
Thermistor
Thermistor
Polymer Resistance
Thermistor
Polymer Resistance
Thermocouple: Type K, J, T, E, S, R (*1)
Measurement Units
°C,°F
°C,°F
%RH
°C,°F
%RH
°C,°F
Measure-
ment Range
Internal Sensor
-10 to 60°C (*2)
-
-
-
-
-
External Sensor
-40 to 110°C (Supplied Sensor)
-60 to 155°C (Optional
Sensor:Fluoropolymer Coated
Type)
0 to 55°C
10 to 95 %RH
-25 to 70°C
0 to 99 %RH (*3)
Type K : -199 to 1370°C
Type J : -199 to 1200°C
Type T : -199 to 400°C
Type E : -199 to 1000°C
Type S : -50 to 1760°C
Type R : -50 to 1760°C
Accuracy
Avg. ±0.3°C
at -20 to 80°C
Avg. ±0.5°C
at -40 to -20°C, 80 to 110°C
±0.5°C
±5 %RH
at 25°C, 50 %RH
±0.3°C
at 10 to 40°C
±0.5°C
at all other temperatures
±2.5 %RH
at 15 to 35°C, 30 to 80 %RH
Thermocouple Measurement
(Sensor inaccuracies not included)
Type K, J, T, E : ±(0.5°C + 0.3 % of reading)
at -100°C or above
Type S, R : ±(1.5°C + 0.3 % of reading)
at 100°C or above
Cold Junction Compensation
±0.5°C
at 10 to 40°C
±0.8°C
other temperatures within the operating environment of
the logger
Measurement Resolution
0.1°C
0.1°C
1 %RH
0.1°C
0.1 %RH
Type K, J, T, E: 0.1°C, Type S, R: Approx. 0.2°C
Responsiveness
Thermal Time Constant:
Approx. 75 sec.
Response Time (90%):
Approx. 190 sec.
Response Time (90%): Approx. 7 min.
Response Time (90%): Approx. 7 min.
-
Logging Capacity
8,000 data sets (One data set consists of readings for all channels in that type of unit.)
Recording Interval
Select from 15 choices: 1, 2, 5, 10, 15, 20, 30 sec. or 1, 2, 5, 10, 15, 20, 30, 60 min.
Recording Mode
Endless (Overwrite oldest data when capacity is full) / One Time (Stop recording when capacity is full)
Auto-upload Interval
Select from 15 choices: OFF (No auto-upload), 1, 2, 5, 10, 15, 20, 30 min. or 1, 2, 3, 4, 6, 12, 24 hrs.
Communication Interfaces
Wired LAN Communication
100BASE-TX/10BASE-T (RJ45 Connector)
Protocol : HTTP(*5), DHCP, DNS
USB Communication
USB 2.0 (Mini-B connector)
Wireless LAN Communication
IEEE 802.11b/g/n (2.4GHz only)
Security (*4): WEP (64bit/128bit),
WPA-PSK(TKIP), WPA2-PSK(AES)
WPS 2.0 : Push Button Configuration
Protocol : HTTP(*5), DHCP, DNS
Bluetooth Communication
Bluetooth 4.2 (Bluetooth low energy)
USB Communication
USB 2.0 (Mini-B connector)
Power (*6)
Battery: AA Alkaline LR6 x 2, AA Ni-MH x 2
External: USB Bus 5V 200mA, AC Adaptor AD-05A2 or AD-05C2, PoE IEEE 802.3af
Battery:
AA Alkaline LR6 x 2, AA Ni-MH x 2
External:
USB Bus 5V 200mA,
AC Adaptor AD-05A2 or AD-05C2
Battery Life (*7)
Approx. 10 days to 1.5 years (*9)
Approx. 10 days to 1
year (*9)
Approx. 10 days to 1 year (*8)(*9)
Dimensions
H 58 mm x W 78 mm x D 26 mm
Weight
Approx. 55 g
Operating Environment
Temperature: -10 to 60°C (*10), Humidity: 90 %RH or less (no condensation)
Temperature: -10 to 60°C
Humidity: 90 %RH or less (no con-
densation)
Software
PC Software (Windows)
(*11)
TR7 for Windows, T&D Graph, T&D Data Server
Mobile Application (iOS, Android)
T&D Thermo
(For compatible OS versions, please refer to the Software page of our website at tandd.com/software/)
*1
: Compatible wire sizes are as follows. Single Wire : 0.32 to Ø 0.65 (AWG 28 - 22), Twisted Wire : 0.08 to 0.32 (AWG 28 - 22), 0.12 or more in diameter, Stripping Length : 9 to 10
*2: When Auto Upload is used frequently, the measurement of the internal sensor may rise by around 0.3°C
.
When using external power, the data logger itself generates heat and the internal sensor will report a temperature much higher than ambient; we recommend using an external temperature sensor in this case.
*3
: When continually used in environments with temperatures above 60°C, accuracy of humidity measurements will decrease over time. Also, humidity cannot be measured at temperatures below -20°C.
*4
: If you wish to use the WPS feature, set the security type of the wireless LAN access point to "WPA2-PSK(AES)" or "None".
*5
: HTTP client. Proxy supported.
*6
: When using external power, the internal temperature of the logger rises.
*7
: Battery life is highly dependant on the Auto-upload interval; at 1 min will give 10 days of usage, and at 12 hours or more will yield the maximum lifetime. Other influential factors include LAN environment, ambient temperature,
recording interval, and battery performance. All estimates are based on operations carried out with a new battery and are in no way a guarantee of actual battery life.
*8
: Shows the estimated battery life with Bluetooth ON. It will be 1.2 times longer with Bluetooth OFF.
*9
: Shows the estimated battery life with Auto-Upload ON.
*10
: -10 to 45°C when using external power (TR-7nw only).
*11
: For installation, it is necessary to have Administrator (Computer Administrator) rights.
The specifications listed above are subject to change without notice.
Important Notices and Disclaimers
In order to properly use this product, please carefully read all documents that accompany the product before using.
»
All rights of the attached documents belong to T&D Corporation. It is prohibited to use, duplicate and/or arrange a part or whole of the attached documents without the permission of T&D Corporation.
» Microsoft and Windows are registered trademarks of Microsoft Corporation in the United States and/or other countries.
» Google, Android, and Google Play are trademarks or registered trademarks of Google Inc.
» Apple and App Store are trademarks or registered trademarks of Apple, Inc. in the U.S. and other countries.
»
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by T&D Corporation is under license.
»
All registered trademarks, company names, product names and logos mentioned herein or for products being used are the property of T&D Corporation or of their respective owners.
»
Specifications, design and other contents outlined in the attached documents are subject to change without notice.
» Please follow the safety precautions outlined in the attached documents carefully. We cannot guarantee nor are we responsible for safety if this product is used in any manner other than was intended.
» On-screen messages in the attached documents may vary slightly from the actual messages.
»
Please notify the shop where you purchased this product or T&D Corporation of any mistakes, errors or unclear explanations in the attached documents.
»
T&D Corporation accepts no responsibility for any damage or loss of income caused by the use of our product.
» Accompanying documents cannot be reissued, so please keep them in a safe place.
» Please read the warranty and provisions for free repair carefully.
For product information or questions contact us at:
For product inquiries, please contact your local distributor. Visit T&D Website for the distributors list.
If you can not find a distributor in your area, please contact our main office in Japan or one of our branch
offices in Europe or America.
tandd.com/purchasing
2021.06 16508120020 (4th Edition)