You must use a
500g
calibrated weight
ONLY
to properly recalibrate the scale.
1.
While the scale is off, press and hold the "
ON/OFF
" button and then press the
"
UNIT
" button 3 times. Release all buttons. A number between 40000 and
180000 will display.
2.
Press the “
ZERO/
” button. The display will show zero.
3.
Place a 500g calibrated weight on the scale platform.
4.
Press the “
ZERO/
” button. The display shows “
CAl
” then 500.0 g.
5.
Remove the weight. Calibration is complete.
Note:
If the calibration is unsuccessful, the display will show a number between
4000 and 70000 instead of 500.0 g. If this occurs, repeat the steps above.
CALIBRATION PROCEDURE
Occasionally the scale's accuracy should be verified by measuring a known weight
of the scale's approximate capacity. Tare the scale by pressing “
ZERO/
”, place
the weight on the platform and note the reading. Calibrate the scale if necessary.
CALIBRATION FUNCTION