background image

- 26  -

 

 
AUTO-R 

A touch on the icon sw itches this unique function ON or OFF. When sw itched 
OFF, the amplitude of SQUA RE w ave signals and SINE w ave signals w ill be as 
assigned during the setup for all frequencies of a program. 

 

When sw itched ON, the microcontroller calculates the signal strength of 
SQUARE w aves on the base of each frequency value follow ing a physiological 
algorithm. The results override the original amplitude settings. 
If AUTO-R is enabled, OSCA is automatically set to OFF. 

 
GLOBAL TIMER 

A touch on the icon sw itches GLOBAL TIMER  ON or OFF. When sw itched OFF, 
the timer settings in the “123” menu may be set individually f or each frequency. 
When sw itched ON, the timer settings in the “123”  menu is the same for all 50 
frequencies of the program. The global timer becomes active, w hen the timer 
setting in the “123”  menu is  modified. 

 
GLOBAL WOBBLE 

GLOBAL WOBBLE ON or OFF. If enabled the w obble settings w ill be incurred for 
all frequencies of a program. 

 
GLOBAL ENV ELOPE 

If GLOBAL ENVELOPE is activated, and the signal form is SQUARE or SINE, 
the amplitude w ill slow ly oscillate betw een 0 and the value assigned to the 
frequency. The background color of the button changes to green w hen activated. 

 
POWER SAVE 

If enabled, the backlight turns to low est possible setting after 30 seconds. Any 
touch on the screen sets the backlight to its preset value. The pow er save 
function saves battery charge during an application so that it is possible to 
ensure a longer run time.

 

 
LOOP   

LOOP ON or  OFF. If enabled, a program started from „123“ or „A BC“ w ill loop 
forever. The only w ay to stop it is w ith the STOP icon. 

 
AUTO OFF 

If the text "OFF"  is visible inside this button, the AUTO POWER OFF function is 
disabled. When touching the button, the number of minutes af ter w hich the 
device sw itches off automatically w ill increase in one minute steps. The 
automatic pow er off function w ill be restarted if the device is operated, a program 
is running or the device is under softw are control. The remaining time is be ing 
displayed in seconds on left/bottom of the screen. 

 
PROTECT  

Button to avoid accidental modification of frequencies in protocols. When 
activated (green), it is not possible to modify frequency values in programs or 
protocols.

 

 
 

Battery history indication 

 

 

 

 

A blue field w ith green vertical lines is displayed on the top of the screen. If the color of  the lines turns yellow  
or red, the battery charge is poor and it is recommended to use the charger for recharge of the battery. The 
distance betw een tw o vertical gray lines indicates a slot of one hour.  
 
A fully charged battery should be good for 3.5 to 5.5 hours of  independent operation.  
 
The charge time for a fully discharged battery is approx. 6 hours. 

 

 

Summary of Contents for F-SCAN5

Page 1: ...uency Synthesizer F SCAN5 Firmware Version FS5 V1 0x INSTRUCTION MANUAL This symbol identifies the equipment as type B Attention CONSULT ACCOMPANYING DOCUMENTS Ordering information F SCAN5 Product id...

Page 2: ...frequency value 13 The properties of frequencies 13 The menu DUO 14 How to start pause or stop a program 16 The menu SWEEP 17 The menu DIRP 18 How to prepare for a DIRP and run it 20 EAP ELECTRO ACUPU...

Page 3: ...ith 4 generators Special accessories like the UV Flat Magnetic Coil Adapter can be connected to the Pow erPort w ithout a separate amplifier The intensity of the sine w ave and square w ave signals d...

Page 4: ...a procedure that has been developed by TB Electronics This feature can be used to adjust output signal to the application in real time The very sensitive measuring method that is used in this procedu...

Page 5: ...ith instruction manual FTB135 Wall charger FTB224 Stainless steel electrodes FTB202A Application cable for electrodes FTB308 5 sets of self adhesive pad electrodes FTB041 DIRP sensor FTB303 USB cable...

Page 6: ...6 Elements for operation A touch on this icon switches the unit OFF CON2 WIDE output port 15MHz CON2 WIDE Mini USB connector CON1 Multi signal output port CON3 PowerPort ON switch Slot for ChipCard...

Page 7: ...uctivity measurement The channels that are currently in use are indicated in the DUO mode menu Signal forms delivered to Multi signal output port SINE DC OFFSET 10Vpp Vpp SQUARE 0 symmetrical Full Wav...

Page 8: ...e connected Connector 3 5mm CON2 At this connector is the WIDE output available w ith a bandw idth up to 15 MHz Standard application parts may be connected Connector 2 5mm on CON3 Connector for the Po...

Page 9: ...isplayed means SERIAL Serial number of device ONS Number of uses Ds Number of DIRPs FLASH ID identifies the type of internal memory FFFFFFFF w ould indicate a memory defect The device must then be che...

Page 10: ...igned to each frequency value Run time TIME Signal form Auto Sine Square Half Wave Square Full Wave SMA RT and OFF WOBBLE Output sw ings around the target frequency ENV ELOP Amplitude oscillates betw...

Page 11: ...ith 03 00 3 minutes and 0 seconds the signal form selected AUTO the icons WOBBLE and ENV ELOP remain inactive their background is w hite How to select a program from the menu 123 A touch on the progr...

Page 12: ...O copies the selected program to the DUO section so that a standard single channel program can also be used as DUO TRIPLE or QUA D application To close the copy paste cut sub men press the X icon top...

Page 13: ...time The factory setting of the runtime per frequency is 3 00 minutes It can be changed after a touch on the above icon and the input of the minutes and seconds to be applied A touch on the ENTER key...

Page 14: ...image indicates four channels and the red black green yellow and blue output plugs on the screen show where to connect the application parts NOTE the blue plug is used for conductivity measurement and...

Page 15: ...d proceed w ith the intended function COPY copies the entry into the paste buffer PASTE copies the contents of the paste buffer into a selected memory position CUT deletes the entry COPY TO 123 copies...

Page 16: ...ude slider and current indicator are displayed differently The amplitude cannot be modified any more As long as the current measuring system does not see a valid current flow ing the amplitude is set...

Page 17: ...al form is AUTO w hich means SQUARE DC OFFSET through 65000 Hz and SINE above that value The amplitude is as selected in menu 123 If OSCA is set the amplitude is set to 60 To modify the amplitude OSCA...

Page 18: ...king on the right side of the screen indicates the output that has to be used with the finger sensor FTB303 The analysis is based on tw o readings 1 MV Measured Value of a resonance 2 CV Conductivity...

Page 19: ...m the remaining graph You can use the arrows to load a program Touch to move w ithin the graph in the direction of Fmin to read the CV MV and CL of individual resonances Touch to move w ithin the grap...

Page 20: ...d black and blue circles 2 Connect the red wire to the cylinder Hold the cylinder in the left hand 3 Touch the button START to start the analysis 4 Wait until the graph line reaches Fmax and completes...

Page 21: ...ect the EA P cable to the sensor port of the F SCA N5 on the left side of the device Connect the EA P Stylus to the blue lead of the EAP cable Connect a handheld or foot electrodew ith the red lead of...

Page 22: ...ime The measured value show n under the table block must have returned to 00 before the next point can be measured All measured values betw een 45 and 55 are considered to be w ithin the NORMAL RANGE...

Page 23: ...hogen list w ith associated frequencies The list is memorized in a nonvolatile memory and simple procedures for computer experts to dow nload updated or individual lists exist The idea of SUB SCAN is...

Page 24: ...ttons w ill have no effect All other buttons can be used as described in the DIRP chapter After a SUB SCA N procedure also know n as selective DIRP all frequencies that belong to a group are filtered...

Page 25: ...w ords are indicated in the selected language For other languages the ABC list entries are also translated EA P SENSE set this slider to the desired sensitivity The standard setting is 20 6 The value...

Page 26: ...to low est possible setting after 30 seconds Any touch on the screen sets the backlight to its preset value The pow er save function saves battery charge during an application so that it is possible...

Page 27: ...ets previously stored in program positions in the F SCAN5 w ill be lost Touch the third line to store the frequency values of the program position presently open in the F SCAN5 on the ChipCard for use...

Page 28: ...GRA MM 1 10 100 1000 10000 PROGRA MM 2 20 30 40 50 60 100 The program name maximum of 20 characters letters of the alphabet only must be set in parenthesis It is follow ed by 1 to 30 frequency values...

Page 29: ...ments to frequencies offered there are applicable The altered program can thus be transferred to other F SCAN models w ith the help of the ChipCard How to start pause or stop a program Touch the icon...

Page 30: ...wave Amplitude 0 to 24Vpp Square signal per manent positive DC OFFSET Amplitude 0 to 12Vpp Frequency range Sine 1Hz to 3 MHz Square guaranteed to100 kHz See list below for typical dependency of ampli...

Page 31: ...rking F SCAN4 Frequency Synthesizer in alleiniger Verantwortung auf dass sich diese Erkl rung bezieht mit der folgenden Norm bereinstimmt declare under our sole responsibility to which this declaratio...

Reviews: