Local Electrical Loopback Bench Test

Purpose:

This test loops the incoming RS-232, RS-422 or RS-485 electrical signal back to the
output terminal blocks in order to verify the cable connections, the electrical interface
driver, signal polarity (for RS-422/RS-485) and the receiver's Integrated Circuitry.

Equipment
Requirements:

One (1) Bit Error Rate Test (BERT) Set with appropriate interface module.

Procedure:

Set up the bench test as shown in Figure 15.  Set 

SW1 

(LOCLB) to the Right position.

Connect a BERT tester to the IN(-) and IN(+) terminal blocks.  Set the tester up as a
DTE device if  RS-530 or RS-449 interface is used.   "TD" and "RD"  LEDs should light
on the TC1540, indicating the status of the looped signal.  The BERT tester should
indicate a "SYNC" signal.  This test should be performed on each individual unit in the
application.

Figure 17.  Local Loopback Bench Test Connection Diagram

TC1540

SW1 to right

Local loopback

RS-232,RS-422,RS-485

to/from TTL conversion 

To BERT tester

Connect TC1540 to RS-530 DTE device

RS-530

DTE DEVICE

(DB25)

TC1540

Tx-

2

Rx-

3

Tx+ 14

Rx+ 16

IN-

OUT-

IN+

OUT+

SIG
GND

SIG

GND

7

Connect TC1540 to RS-232 DTE device

RS-232

DTE DEVICE

(DB25)

TC1540

Tx-

2

Rx-

3

IN-

OUT-

SIG
GND

SIG

GND

7

Connect TC1540 to RS-449 DTE device

RS-449

DTE DEVICE

(DB37)

TC1540

Tx-

4

Rx-

6

Tx+ 22

Rx+ 24

IN-

OUT-

IN+

OUT+

SIG
GND

SIG

GND

19

Summary of Contents for TC1540

Page 1: ...iven or implied that this document is error free or accurate with regard to any specifica tion TC Communications Inc reserves the right to change or modify the contents of this manual at any time with...

Page 2: ...ion 11 Unpacking the Unit 11 Equipment Location 11 Installation Procedure Summary 11 Installation Example for 2 wire RS 485 application 12 Chapter 3 Troubleshooting 14 General 14 Power On LEDs Flash 1...

Page 3: ...shooting Optic options include 850nm and 1310nm Multimode wavelengths 1310nm or 1550nm Single Mode wavelengths The optional optic redundancy feature can provide maximum reliability for crucial applica...

Page 4: ...lock RS 422 485 output 2 wire RS 485 terminal block Signal Ground DIP Switches for Diagnostics and Application Setup LED Indicators for Electrical Optical Input Output Status Optic Transmitter TxA Opt...

Page 5: ...is is idle state Input Off RS422 Voltage on terminal terminal s voltage RS232 Negative input voltage Logic 0 or Space RS422 Voltage on terminal terminal s voltage RS232 Positive input voltage O n Opti...

Page 6: ...w both channels are in the receiving or transmitting mode at all times The RS 485 Driver Receiver transition time is determined by three DIP switches on the TC1540 The local unit s RS 485 transceiver...

Page 7: ...to an echo effect on the electrical connection With the addition of a termination resistor at the beginning or end of the electrical bus this echo effect is greatly reduced The termination resistors a...

Page 8: ...d Pins OUT OUT are outputs or input output for 2 Wire RS 485 Figures 9 2 wire RS 485 Connection Logic Diagram RS 422 RS 485 4 Wire Termination Connection Transition Control Rcv User s Device W1 Termin...

Page 9: ...9 TC1540DUser sManual Rev 1 2 Rs232 Drv Rcv TC1540 To From Fiber User s Device Tx Rx IN OUT SIG GND SIG GND Figures 10 2 wire RS 232 Connection Logic Diagram...

Page 10: ...Launch Power 20dBm 850nm 1310nm MM 62 5 125 m 16dBm 1310nm Single Mode 9 125 m Receiver PINDiode typicalSensitivity 36dBm 850nm 1310nm MM 62 5 125 m 36dBm 1310nmSingleMode 9 125 m Launchpower sensitiv...

Page 11: ...may be reversed Reversed power polarity will not cause any damage to the unit B Connect fiber optic cables between the local remote units as shown in Figure 12 the local unit s optic Tx connects to t...

Page 12: ...s half duplex the polling device and responding RTU take turns transmitting Electrical Interface 2 wireRS 485 Data Rate 9600Baud Async Optic Redundancy Utilized DIP Switch settings SW1 SW2 SW3 SW4 SW5...

Page 13: ...13 TC1540DUser sManual Rev 1 2 Figure 12 4 Wire RS 485 Installation Diagram 4 wire RS 485 Quick Installation local remote Units...

Page 14: ...If Multimode cable is used in a Single Mode application the test results could be erroneous and confusing Calculating the Fiber Optic Loss Budget The fiber optic link and or connectors are frequently...

Page 15: ...Is the unit s Power LED Lit Conduct a Remote Loopback Test Conduct a Remote Loopback Test from Opposite Direction Yes Does the TD LED Light when Transmitting Check Pin Assignments Check Term ination R...

Page 16: ...on a TC1540 unit with a 2 wire RS 485 interface due to its half duplex operation however the Signal Generator feature can be used to facilitate troubleshooting The DIP switch illustrations on each uni...

Page 17: ...om optic Tx to optic Rx Set all the DIP switchs to the Left position Connect a BERT tester to the TC1540 according to Fig 15 Set the tester up as a DTE device The BERT tester should indicate a SYNC si...

Page 18: ...s Set the tester up as a DTE device if RS 530 or RS 449 interface is used TD and RD LEDs should light on the TC1540 indicating the status of the looped signal The BERT tester should indicate a SYNC si...

Page 19: ...module At least two 2 optical jumper cables patch cords with appropriate connectors Procedure Set up the bench test as shown in Figure 16 On the remote unit set SW2 RMTLB to the Right position and con...

Page 20: ...RS 232 RS 422 or 2 or 4 wire RS 485 System Bit Error Rate 1 in 109 or better Visual Indicators POWER O SYNC TD RD COLL DRVMD RS485 Power Source standard 12VDC 250mA optional 24VDC 125mA optional optio...

Reviews: