TDI-Dynaload Division
WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
402828, Rev. B1
Page 65
FQ Frequency
Description
This command is used to set the Frequency in Hertz for square wave
pulsing. The Pulsing Base Time (T1) and Pulsing Peak Time (T2)
will be adjusted accordingly based upon the defined duty cycle.
Command Syntax
FQ <NR2>
Example
FQ 1000.0
This will set the frequency to 1000Hz. If the duty cycle was 30%,
T2 will be set to 300µs and T1 will be set to 700µs.
Query Syntax
FQ?
Returned Parameters TEXT ON: <NR2> Hz
TEXT OFF: <NR2>
I?
Read Current
Description
This query returns the actual current reading in Amps passing
through the load.
Command Syntax
NONE
Example
N/A
Query Syntax
I?
Returned Parameters TEXT ON: <NR2> amps
TEXT OFF: <NR2>