53
Output current Recommended wires
(mm²)
Recommended
wires (AWG)
Recommended lugs/Connector
0A - 10A
2
14AWG
Phoenix contact GIC-2.5/4-ST-7,62.
(Supplied in Accessories kit)
10A - 20A
4
10 AWG
Panduit LCA10-14-L OR equivalent.
20A - 40A
6
8 AWG
Panduit LCA8-14-L OR equivalent.
40A - 80A
16
4 AWG
Panduit LCA4-14-L OR equivalent.
80A - 160A
16
4 AWG
Panduit LCA4-14-L OR equivalent.
Connect two 4 AWG cable in parallel
per terminal.
Table 4-2: Recommended Cable Size for Output Connection
4.9.3
Bus bar Cover Limitation
Bus bar cover has a limited hole diameter. Choose load wires with insulation to fit into output protection
cover. Refer to Table 4-2 for the recommended cable. Wires insulation should fit into the specifications
in Figure 4-18.
Figure 4-18: Bus bar Cover Wire Diameter Limitation
4.9.4
Wire Termination
The wires should be properly terminated with terminals securely attached. DO NOT use non-
terminated wires for load connection at the power supply.
4.9.5
Noise and Impedance Effects
To minimize the noise pickup or radiation, the load wires and remote sense wires should be twisted pairs
to the shortest possible length. Shielding of sense leads may be necessary in high noise
environments. Where shielding is used, connect the shield to the chassis via a rear panel Ground
screw. Even if noise is not a concern, the load and remote sense wires should be twisted-pairs to
reduce coupling, which might impact the stability of power supply. The sense leads should be
separated from the power leads.