83550130 Revision A 

 

2.  Interface setup information 

For normal operation, only the GPIB primary address should be adjusted using switches 1 
through 5 if required. The other switches, 6 through 8 remain in the OFF position. 

2.1. 

GPIB primary address (PA) 

The GPIB primary address (PA) enables identification of all units connected to an IEEE-488 
bus system, and each unit connected to the bus must be assigned a unique PA. 

The Host PC usually has PA=0 and the connected units usually have addresses from 4 
upwards. Normally PHV power supplies are factory configured with the PA set to 8. 

The user can change the factory set PA via a DIP switch located on the IEEE-488 interface 
converter on the rear panel of the unit (see the figure on page 3). It is not necessary to 
open the power supply. 

After changing a configuration switch, the power supply should be turned off for 5 seconds 
and turned on again. 

To change the GPIB PA, the user should only change switches 1 through 5. The unit is 
factory set with switch 4 in the ON position and switches 1, 2, 3, and 5 are in the OFF 
position (PA=8).  

The switch settings follow the binary system: 

Switch 1 has value 1 
Switch 2 has value 2 
Switch 3 has value 4 
Switch 4 has value 8 
Switch 5 has value 16 

Example: To set the GPIB PA to 9, Switches 1 and 4 are ON, switches 2, 3, and 5 are 
OFF. To set the GPIB PA to 20, Switches 3 and 5 are ON, switches 1, 2, and 4 are OFF. 

2.2. 

Other DIP switch settings 

DIP switches 6, 7, and 8 are special functions and are factory set to the OFF position and 
the settings should not be changed. Contact the factory for further information. 