On the lower left corner of the BioSigRZ screen you can set the
Acquisition Filters
. In most
cases the best data is found with the following settings:
Once the settings have been edited, we recommend saving the resulting
fi
le with a new
name, something like "Template_ToneABR.acf" because this
fi
le should be used going
forward as a template for all Tone ABRs for this rig.
Wrapping up
Before running any subjects with any ABR system, we recommend running a
. Only
when the saline test looks good should you proceed to test with a subject.
2.
Acquisition Filtering
HP = 300Hz
LP = 3kHz
NT = 50Hz
The NT
fi
lter should match the frequency of the mains power coming from your wall. You will see papers from
around the world where this setting will vary. If the RZ6 is moved to a new location, check to be sure the NT
fi
lter is set appropriately based on the frequency of the power being used.
Note