Temperature sensor failures will be displayed as ???.?? in the Status tab. If you encounter this issue, contact TDT.
Typical fan speed rates should be 1500 RPM and 3500 RPM under heavy processing loads. Fan failures will be
displayed as ????? in the Status tab. If you encounter this issue, contact TDT.
Unformatted storage arrays will cause an Array not found status to be displayed. This may also be caused by disk
drive failures within the RS4. You may attempt to reformat the storage array. See
information on reformatting. If reformatting is not desired, contact TDT.
Communication errors are compiled per recording session for currently installed streaming ports and will indicate
if a streaming port had a communication failure at some point during the session. Data recorded during the
session may be invalid. Communication errors may result from wiring errors between the RZ2 and RS4. Cycling
power on the RZ2(s) may
fi
x the issue. Refer to the See RS4 to RZ2 Connection Diagram. for a proper wiring
example. If the wiring is correct this may indicate a bad
fi
ber optic cable that will need to be replaced.
Status Tab Errors
RS4 Data Streamer | 34