21
5. Connector specifications
5-3. External trigger signal input
(TRIG IN)
Function
When using an external contact signal as a trigger to
start and stop recording, input the trigger signal here.
Changing from H to L starts recording.
Changing from L to H stops recording.
External triggers must be turned on using the LXK Navi
or remote control (optional) trigger settings screen.
Input circuit format
Input format
Input voltage range: 0–5 V
L level: 0.4 V or less
H level: open or 2 V or more
Connected type
BNC
Connection examples
Recording starts when the external trigger signal input
reaches the L level. Use contact and non-contact (tran-
sistor or TTL open collector) to achieve L level.
,1
Contact signal connection example
Transistor connection example
TTL open collector example
ON when IN is H
5-4. DIGITAL CONTROL input/output
Function
Use to control recording and playback with contact sig-
nals and to connect a remote control unit (option).
Contact input
REC_FWD, REC, FWD, STOP, PAUSE, MARK
Input format
Input voltage range: 0–5 V
L level: 0.4 V or less
H level: open or 2 V or more
Pulse width: 100 msec or more
Status output
REC, FWD, STOP, PAUSE
Output format
Maximum applied voltage: +5 V
Open drain, 8mA maximum sync current
Connected type
Angled, half-pitch, 36-pin (Hirose DX10A-36S)
Continued on the next page