6-4. TEDS calibration
A TEDS sensor has calibration information including the rated
output and capacity stored in its internal memory.
Using TEDS calibration, you can load the calibration
information and automatically register it as calibration values.
Load TEDS data and
modify loaded data
Zero-point calibration
1. To load the TEDS data;
press TEDS, and then press ENTER.
The rated output (mV/V) you loaded appears with the least
significant digit blinking.
If you want to change the displayed data, you can
change the blinking digit of the value by pressing
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
CAUTION
If the bridge voltage setting is higher than the maximum
excitation voltage recorded in the TEDS memory when TEDS
data is loaded, the bridge voltage setting is automatically
changed to a setting equal to or lower than the maximum
excitation voltage.
2. To save the rated output or the change you made,
press ENTER.
The rated capacity appears with the least significant digit
blinking.
3. To set the rated capacity;
If you want to change the displayed data, you can
change the blinking digit of the value by pressing
5
or
b
.
To move to the previous or next digit, press
g
or
t
.
To change the digit, press:
To move to the previous
or next digit, press:
To change the decimal point position, blink the digit
you want to the left of the decimal point by pressing
g
or
t
, and then press FNC.
NOTE
The position of the decimal point you set in this step is used
for readings on the TD-260T.
4. To save the rated capacity or the change you made,
press ENTER.
CAUTION
If you use the D/A option, the rated capacity is set as the D/
A full scale value.