5235
TECSource
User’s Manual · Page 11
Because
5235 TECSource
uses a digital PID loop, it will always stabilize at the
set point, and you will never have a temperature offset, which is different that
most other temperature controllers.
Using Limits
The
TECSource
supports three different types of limits: temperature, sensor,
and current. There is both a temperature high and temperature low limit, and if
the actual temperature exceeds either of these limits, the output will be shut off.
Likewise, both high and low limits exist for the sensor (R limits), and the output
will be shut off if the sensor measurement exceeds either of these limits.