Settings and Menus
All parameters of the
TECSource
can be viewed and changed within the menu.
The menus are constructed with the most used parameters first. To change any
setting, press
Menu
to enter the menu then rotate the knob to select the
parameter to change. Press the knob to begin changing the value. As a visual
indication that you are in edit mode, you will see an asterisk appear next to the
value. Once you have made your change, press the knob or
Menu
button to
store the value. Pressing the
Menu
button will store and exit the menu, while
pressing the knob will store the value but leave you in the menu to make
additional changes.
Some settings are contained inside a sub menu, such as communications
settings. To access the sub menu, simply press the knob to enter the sub menu
when its name is displayed.
Below is a complete list of available settings:
Menu
Description
Factory
Default
Root Menu
Main Menu
Mode
This set the operating mode (
T Mode
,
R
Mode
, or
Ite Mode
) of the temperature
controller.
T Mode
Mount
Specify the mount connected to the
TECSource
. If using a
LaserMount
, select
the appropriate model. Otherwise, select
User Defined
.
User
Defined
I Lim
I Lim
sets the current limit of the temperature
controller. The limit should be set to a value
that is suitable for your Peltier device.
3A
Gain
Gain
controls the response of the
temperature controller. A higher gain value
will cause the controller to respond more
quickly to the difference between the set
point and the actual temperature, while a
lower value will cause it to respond more
slowly. Select
PID
for direct access to the PID
parameters. Read more on setting gain
below.
30
PID P
The proportional term of the PID loop. Will
only be available if
Gain
is set to
PID
.
1