5. Instrument Features
The table below lists the technical specifications for the instrument.
Parameters Characteristics
Measurement Software
controlled
Interface
Serial interface: RS 232
Filter handling
External filter exchange
Up to 4 excitation filters and 4 emission filters in
separate filter slides.
Up to 8 excitation and emission filter slides can be
defined when using software.
Plate definition
Via scanning software
Plate shaking
Orbital or linear shaking - both offer 3 speeds
Light source
High energy Xenon flash lamp
Optics
PMMA fibers and lenses
Detector
Side window, low dark current photomultiplier tube
Power supply
Auto-sensing:100-120V / 220-240V
50/60 Hz
Power consumption
200 VA
Main fuse
T 2A / 250 (slow blow)
Physical
Outer dimensions
Width: 380 mm
Depth: 415 mm
Height: 170 mm
Weight 18.8
kg
Environmental
Ambient temperature
Operation
15°C - 30°C
59°F - 86°F
Non-operation
-20°C - +60°C
-4°F - +140°F
Relative humidity
Operation
90% non-condensing
Overvoltage category
II
Pollution degree
2
Usage Commercial
Noise level
< 70 dBA
Method of disposal
Electronic waste
(infectious waste)
5-6
Operating Manual for GENios, GENios FL, and GENios
Plus
No: I 112 904 Rev No: 1.1
June 2002