5. Programming Procedure
2022-01
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
41
5.2.2
Define/Edit a Program (Define/Edit Menu)
Press - or + to select a program position in which a new
program will be defined or select an already existing
program to edit.
Enter the required name for the program (up to eight
characters). Select a character by moving it with the cursor
keys between the commas and pressing the return key
Select , to delete a character. Select and press the
return key to save the program name and continue to
mode selection.
DEFINE EDIT
< > EXIT OK
PROG1:
- + EXIT OK
PLATE MODE
- + EXIT OK
Press - or + to select between plate or strip mode.
Plate 1, GRE96ft
- + EXIT OK
Press - or + to select other plate types.
NO. OF CYCLES 1
- + EXIT OK
Press - or + to select the number of cycles
(cycles can be repeated up to 9 times).
C1 P1 ASP
- + EXIT OK
Press - or + to select the process step (ASP, DISP, WASH,
SOAK, USER PR.). Press OK to define the selected
process step. Once the process step has been defined the
display appears with the step position incremented (e.g. P1
becomes P2).
FINAL ASP?
NO YES
Select - or + to define the next process step.
Press OK to define the selected process step.
Add a final aspiration step if necessary.
The parameters for the final aspirate step can be selected.
C1 P2 ASP
- + EXIT OK
C2 CYCLE
- + EXIT OK
To add another cycle, press + until CX CYCLE is displayed
(X is the number of the next cycle). Press OK to define the
number of cycles.
PROGRAM END?
- + EXIT OK
Press
– or + and press OK to select PROGRAM END
to finish defining the program.
Once the program has been defined,
the Start menu appears again.
Press OK to define or edit a program on the instrument.
NAME:EXAMPLE
567890, , ABCDE
START
< > EXIT OK