6-95
6.4.10 The Transport Unlock Wizard
General
The
Transport Unlock
wizard is designed to quickly guide the user through the
transportation unlock procedure.
If the user tries to connect to a locked instrument, or by selecting
Instrument
-
Connect …
from
LS Scanner Shell
menu, the
Transport Unlock
wizard is
displayed.
•
The
Start Remove Transport Lock
window gives a short overview of the
main focus of the wizard.
•
In the
Remove Absorber Pad
window the user is asked to remove the
shock absorber pad from the instrument.
•
In the
Reset Transport Lock
window, lock status is removed and
transports are initialized.
•
In the
End Remove Transport Lock
window, the transport unlock
procedure is completed.
Measurements are possible only after the transport lock has been successfully
removed.
The Start Remove Transport Lock window
The
Start Remove Transport Lock
screen gives a short overview of the main
focus of the wizard. A static text displays information about starting the
Remove
Transport Lock
procedure. This is a purely informational window and no actions
must be carried out.
The
Start Remove Transport Lock
window:
The
Next
button guides the user to the next screen of the wizard: the
Remove
Absorber Pad
window.