3. Installation
2004-11
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
3-9
3.4 Power
Requirements
The instrument is auto sensing and it is, therefore, unnecessary to make any
changes to the voltage range.
Check the voltage specifications on the rear panel of the instrument to ensure
that the voltage supplied to the instrument is correct according to the
specification.
The voltage range is 100-120 / 220-240 V.
If the voltage is not correct, please contact your distributor.
Ensure that a fuse of the correct type and rating is fitted.
Fuse = T 2.0 A / 250 V (slow blow) UL recognized
WARNING
RISK OF FIRE!
REPLACE ONLY WITH FUSES
OF THE SAME TYPE AND RATING
.
Caution
When installing or uninstalling the instrument ensure that the instrument
and the computer are both switched off and disconnected from the main
power supply before the RS 232 interface cable or the special connector
cable is connected or removed.
Caution
If the instructions given in this manual are not correctly carried out, the
instrument May either become damaged or May no Longer be Able to
perform Its Procedures correctly and the Accuracy of the instrument
can no longer be guaranteed.