6. LSServer, Scanner Shell and Download
2004-11
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
6-57
Filename Format
Filename
Format
group box
The
Filename Format
group box contains the following elements:
•
The
Barcode
check box is automatically selected, if a barcode reader is
connected to the instrument and is enabled for use. If barcode reading fails,
barcode is set to “xxx”.
•
The
Wavelength
check box must be selected, if the excitation wavelength
is to be part of the image filename.
•
The
Position
check box must be selected, if the slide position is to be part
of the image filename.
The
Counter
check box is automatically selected and cannot be reset by the
user.
When using standard name (
Use standard name
check box is selected) the
filename format is fixed to:
•
Barcode - excitation wavelength - slide position - current date - image
counter, if a barcode reader is installed and the barcode option is selected.
•
Excitation wavelength - slide position - current date - image counter, if no
barcode reader is installed.
If no check boxes are enabled, the
Filename Format
cannot be changed.
Clear
Use standard name
to enable the following check boxes / option buttons
and the
Date Format
group box:
•
Wavelength
check box
•
Position
check box
Filename Format
is user-definable.
Date Format
Date
Format
group box
•
The
Date Format
is automatically selected and cannot be reset by the user,
but the type of date format can be selected:
▪
DDMMYY - Day Month Year
▪
YYMMDD - Year Month Day
Stitching
Stitch
resulting
images
check box
Select the
Stitch resulting images
check box to combine all scanned slides
together into one image.
The filenames used for parts of the stitched images are automatically set to
standard name.
The filename of the stitched image is set to:
Standard name
+ fixed
Filename
Format
, where barcode option and slide position are set to “xxx”.