6. LSServer, Scanner Shell and Download
2004-11
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
6-63
When using a scanlist file the following dialog box will appear after clicking
Start
:
Click
Yes
and the following dialog box will appear:
Click the
Proceed
button to skip the warm up time of the laser. If the
Proceed
button is not clicked the scan starts automatically after the wait time has expired.
Example2:
S,180,0,633,670,0,0
DP,220,190,594,635,488,535
DS,120,100,633,670,488,535
Scan Mode: S ( Single ), DS ( Dual Sequential), DP ( Dual Parallel ), Scan Mode,
Gain Channel1, Gain Channel2, Laser Channel1, Filter Channel1, Laser
Channel2, Filter Channel2
In this example, three different scans are performed:
First a single scan is performed on each scan area of the adapter with a gain of
180, with red laser and red filter.
Next a dual parallel scan is performed with a gain of 220 in the first channel with
orange laser and orange filter and a gain of 190 in the second channel with blue
laser and blue filter.
Then a dual sequential scan is performed with a gain of 120 in the first channel
with red laser and red filter and a gain of 100 in the second channel with blue
laser and blue filter.