3 - Technical Data
Electrical Module Interface
MIO2 Operating Manual, 394934, en, V3.0
3 - 7
3.5 Electrical Module Interface
Supply Ratings
Electrical
Safety
Classification according to EN 61010-1 with regard to electrical safety:
3.6 Requirements
The interface cable is a 9 Pin Sub D CAN-Bus connecting the instrument with the
MIO2.
Warm-up time from
20°C to 45°C
max. 30 min.
Warm-up time from
20°C to 60°C
max. 30 min.
Shaking / rotation fre-
quency
1–8.5 Hz
Linear shaking
max. 2 mm amplitude in Y-dimension
Tab. 3-7
Operating ranges, 4 positions
Tab. 3-8
Operating ranges
Connector
9 pin sub D CAN-Bus
Power consumption
<150W (90W
a)
)
a) Setup in current limitation mode with the Instrument Software 6.0 or later version.
Voltage range
24V DC (±10%)
Fuses
T8A 250V
Tab. 3-9
Electrical specifications (safety)
Overvoltage category
II
Tab. 3-10
Software requirements MIO2
6 Positions
4 Positions
Setup & Service
Instrument Software
V 6.0 or later
V 6.4 or later
Application Software
Freedom EVOlution
V 2.0 or later
-
Application Software
EVOware
V 1.4 SP1 or later
V 1.4 SP1 or later
Application Software
FluentControl
V 1.0 or later
Setup & Service
FluentSetup
V 1.0 or later
Summary of Contents for MIO2
Page 14: ...2 6 MIO2 Operating Manual 394934 en V3 0 2 Safety Decontamination Declaration...
Page 32: ...5 8 MIO2 Operating Manual 394934 en V3 0 5 Installation Incubator Installation...
Page 40: ...7 4 MIO2 Operating Manual 394934 en V3 0 7 Maintenance Special Maintenance Tasks...
Page 44: ...9 2 MIO2 Operating Manual 394934 en V3 0 9 Disposal...
Page 48: ...10 4 MIO2 Operating Manual 394934 en V3 0 10 Customer Support Contacts...
Page 50: ...11 2 MIO2 Operating Manual 394934 en V3 0 11 Index...