7 - 2
MIO2 Operating Manual, 394934, en, V3.0
7 - Maintenance
General
7.3 General
WARNING
Injuries or contamination with hazardous materials possible if the maintenance is
not carried out properly.
Make sure that the instrument worktable and the MIO2 are free from
hazardous materials. Decontaminate them (or have them decontaminated) if
necessary.
Protect your body with adequate protective clothing, goggles and gloves.
Make sure MIO2 has cooled down to room temperature before performing
maintenance tasks.
Make sure the incubator slots are empty before performing maintenance
tasks.
Once installed, the MIO2 is part of the instrument.
Decontaminate and clean the MIO2 when decontaminating and cleaning the
instrument.
For detailed information on maintenance of the instrument, refer to the Operating
and Service Manuals of the instrument. Refer to section
7.4 Special Maintenance Tasks
7.4.1 Incubator Removal
1
Make sure the electrical power of the instrument is switched off.
ATTENTION
Risk of damage to equipment. Incubators that are installed in disorder will lead to
damage of equipment.
Only remove one incubator at the time if multiple incubators are installed to
avoid muddling up.
2
Lift and turn the incubator to remove the D-Sub cable connected to the MIO2.
3
Completely remove the incubator.
Summary of Contents for MIO2
Page 14: ...2 6 MIO2 Operating Manual 394934 en V3 0 2 Safety Decontamination Declaration...
Page 32: ...5 8 MIO2 Operating Manual 394934 en V3 0 5 Installation Incubator Installation...
Page 40: ...7 4 MIO2 Operating Manual 394934 en V3 0 7 Maintenance Special Maintenance Tasks...
Page 44: ...9 2 MIO2 Operating Manual 394934 en V3 0 9 Disposal...
Page 48: ...10 4 MIO2 Operating Manual 394934 en V3 0 10 Customer Support Contacts...
Page 50: ...11 2 MIO2 Operating Manual 394934 en V3 0 11 Index...