ULTRA CLEAN OILLESS COMPRESSOR
INSTALLATION
2
2.
INSTALLATION STEPS
This dental compressor should only be installed by qualified personnel. Should any questions arise
during the installation, call Tech West Technical Support between the hours of 7:00 a.m. to 4:00 p.m.
(Pacific Standard Time).
Place the compressor in a clean, dry, well ventilated area, on a solid, level surface. Consider sound
level and insulate as needed. Be sure that adequate ventilation is available as the compressor is air
cooled. Ambient temperature in the equipment room should be within the temperature range of 40
degrees Fahrenheit minimum to 100 degrees Fahrenheit maximum.
(a) Check the shipping carton for damage. This could detect damage to the unit which might
otherwise be overlooked. Remove cardboard shipping carton.
(b) Remove the Oilless Compressor from its shipping skid. Inspect the unit for damage.
Oilless Compressors are shipped bolted to a pallet. This pallet is intended for shipping only
and should be discarded.
(c) Remove installation kit attached to pallet. It should contain the following:
(4) Isolation Feet
(1) Alternate Air Hookup Hose
(1) 5’ Flexible Air Hose
(d) Install isolation feet on tank legs.
(e) Move compressor into place and level by observing bubble level on compressor platform.
(f) Wiring instructions:
(1) Have all electrical connections made by qualified personnel only. All connections
should be in accordance with local codes.
(2) Use the chart on page 1 to help determine the proper line and breaker size for
the unit that is being installed.
(g) Install the air line from the compressor tank to the building supply.
(h) Install the 1” flex alternate air hose from the compressor to a fresh air supply.