CHAPTER 3 OPERATION
27
△
Note
1. When turning on the instrument, confirm that there is
not any sample or material blocking light inside sample
compartment.
2. Every time instrument is turned on, 1nm slit is used
automatically. Wavelength is calibrated using feature
peak of D2 lamp at 656.1nm.
3. To save initialization time, instrument will not calibrate
wavelength of other slits. Instrument saves wavelength
position of other slits using memory function.
4. If high precision measurement at certain slit level is
required, or wavelength correctness of each slit should
be improved, please set the slit level needed, and
perform wavelength calibration under this slit.
Instrument will save the wavelength position of this slit
level.
5. For D2 lamp and instrument pre-heating, it is suggested
that wavelength calibration should be done 30 minutes
after powering on.
6. As instrument has photomultiplier auto protection
function, correct reading will be obtained 3 seconds
after closing cover of sample compartment.
7. As instrument has photomultiplier auto protection
function, we still suggest that cover of sample
compartment should be closed usually.