CHAPTER 4 MAINTENANCE
34
4.3 Troubleshooting
If there is an abnormality, correct it referring to the following table
4.3.1 Error Information
If failure occurs during operation, error information will be displayed on
instrument.
Symptom
Cause
Solution
ROM
!
ROM broken
Contact to serviceman
RAM
!
RAM broken
Contact to serviceman
656
.
1nm
!
656
.
1nm not found
Check whether sponge in
viewfinder has been taken out,
whether any material in sample
compartment, and do
wavelength calibration again.
Wavelength
drive
!
Wavelength drive error
Contact to serviceman
Baseline
!
Current baseline parameter not
complied with setting parameter
Reset parameters for baseline
calibration.
D2
!
When setting D2 lamp ON or need to
use D2 lamp, lamp is not ON.
Turn on D2 lamp
WI
!
When setting WI lamp ON or need to
use WI lamp, lamp is not ON.
Turn on WI lamp